Preserving image texture while reducing radiation dose with a deep learning image reconstruction algorithm in chest CT: A phantom study |
| |
Affiliation: | 1. IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30029 Nîmes, France;2. Department of Medical Physics, Nîmes University Hospital, 30029 Nîmes Cedex 9, France;1. Department of Diagnostic Radiology, Hiroshima University, 1-2-3 Kasumi, Minami-ku, Hiroshima 734-8551, Japan;2. Canon Medical Research USA, Vernon Hills, IL, United States;3. Canon Medical Systems, Otawara, Tochigi, Japan |
| |
Abstract: | PurposeTo assess whether a deep learning image reconstruction algorithm (TrueFidelity) can preserve the image texture of conventional filtered back projection (FBP) at reduced dose levels attained by ASIR-V in chest CT.MethodsPhantom images were acquired using a clinical chest protocol (7.6 mGy) and two levels of dose reduction (60% and 80%). Images were reconstructed with FBP, ASIR-V (50% and 100% blending) and TrueFidelity (low (DL-L), medium (DL-M) and high (DL-H) strength). Noise (SD), noise power spectrum (NPS) and task-based transfer function (TTF) were calculated. Noise texture was quantitatively compared by computing root-mean-square deviations (RMSD) of NPS with respect to FBP. Four experienced readers performed a contrast-detail evaluation. The dose reducing potential of TrueFidelity compared to ASIR-V was assessed by fitting SD and contrast-detail as a function of dose.ResultsDL-M and DL-H reduced noise and NPS area compared to FBP and 50% ASIR-V, at all dose levels. At 7.6 mGy, NPS of ASIR-V 50/100% was shifted towards lower frequencies (fpeak = 0.22/0.13 mm−1, RMSD = 0.14/0.38), with respect to FBP (fpeak = 0.30 mm−1). Marginal difference was observed for TrueFidelity: fpeak = 0.33/0.30/0.30 mm−1 and RMSD = 0.03/0.04/0.07 for L/M/H strength. Values of TTF50% were independent of DL strength and higher compared to FBP and ASIR-V, at all dose and contrast levels. Contrast-detail was highest for DL-H at all doses. Compared to 50% ASIR-V, DL-H had an estimated dose reducing potential of 50% on average, without impairing noise, texture and detectability.ConclusionsTrueFidelity preserves the image texture of FBP, while outperforming ASIR-V in terms of noise, spatial resolution and detectability at lower doses. |
| |
Keywords: | Computed tomography Deep learning image reconstruction Image quality Dosimetry Iterative reconstruction Contrast-detail evaluation Chest |
本文献已被 ScienceDirect 等数据库收录! |
|