Evaluation of an automatic method for detection of defects in linear and curvilinear ultrasound transducers |
| |
Affiliation: | 1. Department of Radiation Physics, Institute of Clinical Sciences at Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden;2. Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg, Sweden |
| |
Abstract: | PurposeThe high incidence of defective ultrasound transducers in clinical practice has been shown in several studies. Recently, a novel method using only stored images for automatic detection of defective transducers was presented. The method makes it possible to remotely monitor many transducers at the same time and send a notification when a defective transducer is found. The purpose of the present study was to evaluate the novel method and assess how well it performs when compared to an established method as reference.MethodsTo evaluate the novel method, in-air images were collected from 81 transducers in radiologic departments in nine hospitals. Two observers assessed the in-air images and marked the defects. Receiver operating characteristic (ROC)- and alternative free response receiver operating characteristic (AFROC)-curves and their figures of merit (FOM) were calculated for the novel method, using marked defects in the in-air images as reference truth.ResultsThe area under the ROC curve was 0.88 (SD 0.06), and the AFROC FOM was 0.71 (SE 0.07).ConclusionThe result shows that the novel method has a good agreement with the in-air method for detecting defects in ultrasound systems. This indicates that the novel method could be a complement to the normal quality control for early, and automatic detection of defects. |
| |
Keywords: | Transducer Probe Ultrasound |
本文献已被 ScienceDirect 等数据库收录! |
|