首页 | 本学科首页   官方微博 | 高级检索  
     


Compatibility of Stabilized Whole Blood Products with CD4 Technologies and Their Suitability for Quality Assessment Programs
Authors:Tao Ding  Michèle Bergeron  Peggy Seely  Xuefen Yang  Tamsir O. Diallo  Margot Plews  Paul Sandstrom  T. Blake Ball  Adrienne F. A. Meyers
Affiliation:1. National Laboratory for HIV Immunology, Public Health Agency of Canada, Ottawa, Ontario, Canada.; 2. Department of Medical Microbiology, University of Manitoba, Winnipeg, Manitoba, Canada.; 3. Department of Immunology, University of Manitoba, Winnipeg, Manitoba, Canada.; 4. Department of Medical Microbiology, University of Nairobi, Nairobi, Kenya.; 5. National Laboratory for HIV Immunology, Public Health Agency of Canada, Winnipeg, Manitoba, Canada.; University of Cape Town, South Africa,
Abstract:

Background

CD4 T cell enumeration is the most widely used prognostic marker for management of HIV disease. Internal quality control and external quality assessment (EQA) programs are critical to ensure reliability of clinical measurements. The utility of stabilized whole blood products (SWBP) as a test reagent for EQA programs such as Quality Assessment and Standardization for Immunological measures relevant to HIV/AIDS (QASI) program have been demonstrated previously. Since then, several new commercial SWBPs and alternative CD4 enumeration technologies have become available. Seven SWBPs were evaluated on seven different enumeration platforms to determine which product(s) are most suitable for EQA programs that support multiple analytical technologies.

Method

Assessment of SWBPs was based on two criteria: (1) accuracy of CD4 T cell measurements and; (2) stability under sub optimal storage conditions.

Results

Three SWBPs (Multi-Check, StatusFlow and CD4 Count) showed accurate CD4 T-cell absolute count and percentage values across six of the enumeration platforms. All products retain stability up to 18 days at 21–23°C with the exception of Multi-Check-high on FacsCount and Multi-Check-Low and StatusFlow-Low on Pima. One of the products (CD4 Count) retained stability for three days on all platforms tested when stored at 37°C.

Conclusion

This study demonstrated that the characteristics of commercially available SWBPs vary across multiple CD4 platforms. The compatibility of testing panels for EQA programs with multiple analytical platforms needs to be carefully considered, especially in large multiplatform CD4 EQA programs. The selection of a suitable cross-platform SWBP is an increasing challenge as more reagents and platforms are introduced for CD4 T-cell enumeration.