首页 | 本学科首页   官方微博 | 高级检索  
   检索      


A Thermoelectrically Cooled Microtome Table and Knife
Authors:T Rutherford  W S Hardy  P A Isherwood
Institution:  a Unilever Research Laboratory, Bedford, England
Abstract:Thermoelectric cooling units (Frigistor thermoelements) have been used to replace CO2 gas and solid CO2 for microtome stage and knife cooling. These units consist of assemblies of series-connected thermoelements, functioning by the Peltier effect. Cooling is controlled by the direct current supply to the units. Current supply is from a double power pack giving 15 amp at 4.8 v for the knife cooling units, and 15 amp at 1 v for the stage. By varying the current flow, the optimum cutting temperature can be obtained and held indefinitely. An 8-couple Frigistor unit replaces the CO2 stage of a Lipshaw freezing microtome. The stage temperature may be lowered to -36 C in 40-60 sec and at the optimum cutting temperature, 5 μ serial sections of fixed frozen tissue are obtainable. Four 12-couple units are used to cool a 160 mm Jung plane wedge microtome knife fitted to a Reichert sledge microtome, with the stage cooled by one 8-couple unit. The knife temperature can be lowered to -20 C in 5 min; the stage in 1 min. The apparatus has been used to cut a variety of unfixed rat and mouse tissues. The optimum sectioning temperature for such unfixed liver, kidney, spleen, lymph glands, heart, testes, small intestine, pancreas, skin and lung was -20 to -22 C.
Keywords:
本文献已被 InformaWorld 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏