Customized mandibular reconstruction plates improve mechanical performance in a mandibular reconstruction model
Authors:Ralf Gutwald  Raimund Jaeger  Floor M. Lambers
Affiliation:1. Department of Oral and Maxillofacial Surgery, University Hospital of Freiburg, Freiburg, Germany;2. Fraunhofer Institute for Mechanics of Materials IWM, Polymer Tribology &3. Biomedical Materials – Group, Freiburg, Germany;4. Stryker Leibinger GmbH &5. Co. KG, Freiburg, Germany
Abstract:The purpose of this paper was to analyze the biomechanical performance of customized mandibular reconstruction plates with optimized strength. The best locations for increasing bar widths were determined with a sensitivity analysis. Standard and customized plates were mounted on mandible models and mechanically tested. Maximum stress in the plate could be reduced from 573 to 393 MPa (?31%) by increasing bar widths. The median fatigue limit was significantly greater (p < 0.001) for customized plates (650 ± 27 N) than for standard plates (475 ± 27 N). Increasing bar widths at case-specific locations was an effective strategy for increasing plate fatigue performance.
Keywords:Mandibular reconstruction  sensitivity analysis  finite element analysis  fatigue testing  implant