Estimation of blood pressure-related parameters by electrical impedance measurement. |
| |
Authors: | J H Muntinga K R Visser |
| |
Institution: | Department of Medical Physiology, University of Groningen, The Netherlands. |
| |
Abstract: | In 13 healthy volunteers a computerized experimental set-up was used to measure the electrical impedance of the upper arm at changing cuff pressure, together with the finger arterial blood pressure in the contralateral arm. On the basis of a model for the admittance response, the arterial blood volume per centimeter length (1.4 +/- 0.3 ml/cm), the venous blood volume as a percentage of the total blood compartment (49.2 +/- 12.6%), and the total arterial compliance as a function of mean arterial transmural pressure were estimated. The effective physiological arterial compliance amounted to 2.0 +/- 1.3 microliters.mmHg-1.cm-1 and the maximum compliance to 33.4 +/- 12.0 microliters.mmHg-1.cm-1. Additionally, the extravascular fluid volume expelled by the occluding cuff (0.3 +/- 0.3 ml/cm) was estimated. These quantities are closely related to patient-dependent sources of an unreliable blood pressure measurement and vary with changes in cardiovascular function, such as those found in hypertension. Traditionally, a combination of several methods is needed to estimate them. Such methods, however, usually neglect the contribution of extravascular factors. |
| |