High-performance liquid chromatographic method for the simultaneous determination of HIV-1 protease inhibitors indinavir,saquinavir and ritonavir in plasma of patients during highly active antiretroviral therapy |
| |
Affiliation: | 1. Chemistry Department, Faculty of Sciences and Technology, University of Coimbra, 3004-535 Coimbra, Portugal;2. Serviços Farmacêuticos, Centro Hospitalar e Universitário de Coimbra, EPE, 3000-075 Coimbra, Portugal |
| |
Abstract: | A new high-performance liquid chromatographic method for the simultaneous determination of indinavir, saquinavir and ritonavir in human plasma is described. Quantitative recovery following liquid–liquid extraction with diethyl ether from 500 μl of human plasma was achieved. Subsequently, the assay was performed with a linear gradient starting at 67 mM potassium dihydrogenphosphate–acetonitrile (65:35 to 40:60, v/v) as a mobile phase, a Phenomenex C18 column and UV detection at 240 and 258 nm, respectively. Linear standard curves were obtained for concentrations ranging from 75 to 20 000 ng/ml for indinavir, from 10 to 6000 ng/ml for saquinavir, and from 45 to 30 000 ng/ml for ritonavir. The calculated intra- and inter-day coefficients of variation were below 6%. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|