首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Simultaneous enzymatic/electrochemical determination of glucose and L-glutamine in hybridoma media by flow-injection analysis
Authors:Meyerhoff M E  Trojanowicz M  Palsson B O
Institution:Department of Chemistry, University of Michigan, Ann Arbor, MI 48109-1055.
Abstract:A split-stream flow-injection analysis system is described for simultaneous determination of glucose and L-glutamine in serum-free hybridoma bioprocesses media. Amperometric measurement of glucose is based on anodic oxidation of hydrogen peroxide produced by immobilized glucose oxidase within a triple layer membrane of an integrated flow-through glucose-selective biosensor. Determination of L-glutamine is based on quantitating ammonium ions produced in a flow-through enzymes reactor containing immobilized glutaminase enzyme, and subsequent downstream potentiometric detection of these ions by a nonacting-based ion-selective polymer membrane electrode. Endogenous potassium and ammonium ion interference in the L-glutamine determination are eliminated by using a novel in-line tubular cation-exchange membrane unit to exchange these interferent species for cations undetectable by the membrane electrode. The first generation split-steam flow-injections system can assay 12 samples/h using direct injections of 50 muL of media samples, with linear responses to glucose in the range of 0.03 to 30mM, and log-linear response to L-glutamine from 0.1 to 10 mM. (c) 1993 Wiley & Sons, Inc.
Keywords:hybridoma cultivation media  glucose  L-glutamine  flow-injection analysis  biosensors
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏