Flow cytometric prescreening of cervical smears.
Authors:W A Linden  K Ochlich  H Baisch  K U Scholz  H J Mauss  H E Stegner  D S Joshi  C T Wu  I Koprowska  C Nicolini
Abstract:One-parameter (nuclear DNA) and two-parameter (nuclear DNA and protein or cellular light scatter) measurements of cervical smears were performed using an ICP 11 and a cytofluorograf 4800 respectively. A total of about 1000 cases was analyzed. For the estimation of nuclear DNA alone two fluorochromes were tested (ethidium bromide (EB) and mithramycin (MMC)) combined with three different methods of cell preparation. For the two-parameter measurements cells were double stained with EB and fluorescein isothiocyanate (FITC). Red fluorescence (EB) versus green fluorescence (FITC) or red fluorescence versus scatter were recorded. A computer analysis of the one-parameter histograms was performed using discriminant analysis and the results were compared with the cytodiagnosis of microscopic specimens stained with the Papanicolaou technique. The error rates of the flow cytometric (FCM) data were as follows: (a) standard EB staining, 11% false negative, 26% false positive, 6% unsatisfactory results; (b) pepsination of vital cells and EB staining, 12% false negative, 14% false positive and 4% unsatisfactory results; (c) MMC staining, 10% false negative, 65% false positive and 5% unsatisfactory results. Our two-parameter measurements prove that, as confirmed by cell sorting, red fluorescence versus scatter allows separation of at least three subpopulations in most analyzed samples: (a) anucleated cells; (b) leukocytes; and (c) intermediate and superficial cells.