首页 | 本学科首页   官方微博 | 高级检索  
   检索      


2-Propanol in the mobile phase reduces the time of analysis of CLA isomers by silver ion-HPLC
Authors:Katrin Kuhnt  Christian DegenGerhard Jahreis
Institution:Institute of Nutrition, Friedrich Schiller University, Dornburger Str. 24, D-07743 Jena, Germany
Abstract:Individual isomers of octadecadienoic acid (C18:2) with conjugated double bonds (conjugated linoleic acids; CLA) exert different biological activities. Their distribution in food and tissues differs. Therefore, the separation of the various positional and geometric isomers is important. The time of analysis using silver ion-high performance liquid chromatography can extend up to 90 min. The aim of this study was to reduce this time. The time of analysis reduced from ca. 90 min onto 45 to 35 min, respectively, by the addition of 0.05% or 0.1% (v/v) 2-propanol to the mobile phase acetonitrile (0.1%; v/v) and diethyl ether (0.5%; v/v) in n-hexane]. There was no effect on resolution of the 17 individual CLA isomers of the CLA mixture. Regarding the lowest coefficient of variation and an adequate baseline separation the use of 0.05% 2-propanol in the mobile phase is recommended, without any disadvantages and adverse effects on the service life of columns. In conclusion, adding 0.05% or 0.1% 2-propanol to the mobile phase shortens the time of analysis of CLA isomers, saves solvents and reduces costs.
Keywords:Conjugated linoleic acids  Silver ion-high performance liquid chromatography  Time of analysis  Geometric isomers
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏