Comparaison de la dose efficace en TEP/TDM et en TDM
Authors:J Coulot  F MagnierP Chauchat  N GuilabertM Ricard  S Dreuil  J Lumbroso
Institution:a Service de physique médicale, institut de cancérologie Gustave-Roussy, 114, rue Édouard-Vaillant, 94805 Villejuif cedex, France
b Service de médecine nucléaire et cancérologie endocriennne, institut de cancérologie Gustave-Roussy, 114, rue Édouard-Vaillant, 94805 Villejuif cedex, France
Abstract:Hybrid imaging, particularly positron emission tomography (PET) combined with CT has emerged in the field of oncology as a modality of choice. The concomitant realization of a standard CT examination, however, raises the question of the additional dose delivered to the patient. This radiation burden could be avoided by performing a single PET/CT examination with injection of contrast media. To verify the potential dosimetric gain of this strategy, we compared the effective dose associated with each modality in a retrospective cohort of 151 patients, homogeneous in weight and size. The average effective dose for a PET/CT (injection of 5-6 MBq/kg of 18FDG) was 13.5 mSv, the CT dose representing approximately 80% of the PET dose. In our study, the average effective dose in CT thorax/abdomen/pelvis was 21.4 mSv, 60% higher than the PET/CT effective dose.
Keywords:Dose au patient  Scanographie  TEP/TDM