首页 | 本学科首页   官方微博 | 高级检索  
   检索      


96-Well liquid-liquid extraction liquid chromatography-tandem mass spectrometry method for the quantitative determination of ABT-578 in human blood samples
Authors:Ji Qin C  Todd Reimer M  El-Shourbagy Tawakol A
Institution:Abbott Laboratories, Department of Drug Analysis, 100 Abbott Park Road, Abbott Park, IL 60064-6126, USA. qin.ji@abbott.com
Abstract:We report here a quantitative method for the analysis of ABT-578 in human whole blood samples. Sample preparation was achieved by a semi-automated 96-well format liquid-liquid extraction (LLE) method. Aluminum/polypropylene heat seal foil was used to enclose each well of the 96-well plate for the liquid-liquid extraction. A liquid chromatography combined with tandem mass spectrometry (LC-MS/MS) method with pre-column regeneration was developed for the analysis of sample extracts. Selective reaction monitoring (SRM) of the mass transitions m/z 983-935 and m/z 931-883 was employed for the detection of ABT-578 and internal standard, respectively. The ammonium adduct ions M + NH(4)](+) generated from electrospray ionization were monitored as the precursor ions. The assay was validated for a linear dynamic range of 0.20-200.75ng/ml. The correlation coefficient (r) was between 0.9959 and 0.9971. The intra-assay CV (%) was between 1.9 and 13.5% and the inter-assay CV (%) was between 4.7 and 11.3%. The inter-assay mean accuracy was between 86.4 and 102.5% of the theoretical concentrations.
Keywords:ABT-578
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏