Determination of 5-fluorouracil in microvolumes of human plasma by solvent extraction and high-performance liquid chromatography
Authors:Walter J. Loos   Peter de Bruijn   Lia van Zuylen   Jaap Verweij   Kees Nooter   Gerrit Stoter  Alex Sparreboom
Affiliation:Laboratory of Experimental Chemotherapy and Pharmacology, Department of Medical Oncology, Rotterdam Cancer Institute (Daniel den Hoed Kliniek) and University Hospital Rotterdam, P.O. Box 5201, 3008 AE Rotterdam, The Netherlands
Abstract:In the present study, a new reversed-phase HPLC method has been developed and validated for the quantitative determination of 5-fluorouracil (5-FU) in human plasma using only 100-μl samples. The sample extraction and clean-up procedure involved a simple liquid–liquid extraction after addition of 5-chlorouracil (5-CU), used as internal standard, with 5 ml ethyl acetate. Chromatographic separations were performed on an Inertsil ODS-3 column (250×4.6 mm ID; 5 μM particle size), eluted with a mobile phase composed of acidified water (pH 2.0). The column effluent was monitored by UV absorption measurement at a wavelength of 266 nm. The calibration curves were constructed over a range of 0.20–50.0 μM and were fitted by weighted (1/x) linear regression analysis using the ratio of peak heights of 5-FU and 5-CU versus concentrations of the nominal standards. Extraction recoveries over the total range averaged 92 and 93% for 5-FU and 5-CU, respectively. The lower limit of quantitation was established at 0.20 μM (not, vert, similar26 ng/ml), with within-run and between-run precisions of 4.2 and 7.0%, respectively, and an average accuracy of 109.3%. The within-run and between-run precisions at four tested concentrations analyzed in quintuplicate over a time period of four days were <1.4 and <4.4%, respectively. The accuracy at the tested concentrations ranged from 98.4 to 102.3%. Compared to previously described validated analytical methods for 5-FU, our present assay provides equivalent to superior sensitivity using only microvolumes of sample.
Keywords:5-Fluorouracil
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号