首页 | 本学科首页   官方微博 | 高级检索  
     


Reversible surface thiol immobilization of carboxyl group containing haptens to a BIAcore biosensor chip enabling repeated usage of a single sensor surface
Authors:Schlecht Ulrich  Nomura Yoko  Bachmann Till  Karube Isao
Affiliation:Research Center for Advanced Science and Technology, University of Tokyo, 4-6-1 Komaba, Meguro-ku, Tokyo 153-8904, Japan.
Abstract:We describe a reversible immobilization method for carboxyl group containing haptens that makes the repeated usage of a BIAcore biosensor chip possible. Haptens which are immobilized according to the surface thiol method can be removed completely from the sensor surface again by a reducing step. In the first part of our study, analogues of the herbicides 2,4-dichlorophenoxyacetic acid and 2,4,5-trichlorophenoxyacetic acid were immobilized in succession to a biosensor surface of a BIAcore surface plasmon resonance instrument according to the thiol coupling method. Direct kinetic analysis of these ligands to a polyclonal anti-2,4-dichlorophenoxyacetic acid antibody were performed using these biosensor surfaces. In the second part of the study, different amounts of 2,4-dichlorophenoxyacetic acid were sequentially immobilized onto the same biosensor surface in order to generate a calibration plot for 2,4-dichlorophenoxyacetic acid. Using this plot, the quantitative detection of the herbicide down to a concentration of 0.1 microg/mL, the maximum admissible concentration of pesticides in drinking water, is possible.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号