The rationale and a computer evaluation of a gamma irradiation sterilization dose determination method for medical devices using a substerilization incremental dose sterility test protocol
Authors:KW Davis  WE Strawderman  JL Whitby  
Institution:ETHICON, Inc., Route 22 Somerville, New Jersey 08876, USA;Rutgers University, New Brunswick, New Jersey, USA;University Hospital, London, Ontario, Canada
Abstract:The experimental procedure described is designed to allow calculation of the radiation sterilization dose for medical devices to any desired standard of sterility assurance. The procedure makes use of the results of a series of sterility tests on device samples exposed to doses of radiation from 0.2 to 1.8 Mrad in 0.2 Mrad increments. From the sterility test data a 10-2 sterility level dose is determined. A formula is described that allows a value called DS Mrad to be calculated. This is an estimate of the effective radiation resistance of the heterogeneous microbial population remaining in the tail portion of the inactivation curve at the 10-2 dose and above. DS Mrad is used as a D 10 value and is applied, in conjunction with the 10-2 sterility level dose, to an extrapolation factor to estimate a sufficient radiation sterilization dose. A computer simulation of the substerilization process has been carried out. This has allowed an extensive evaluation of the procedure, and the sterilization dose obtained from calculation to be compared with the actual dose required. Good agreement was obtained with most microbial populations examined, but examples of both overdosing and underdosing were found with microbial populations containing a proportion of organisms displaying pronounced shoulder inactivation kinetics. The method allows the radiation sterilization dose to be derived from the natural resistance of the microbial population to gamma sterilization.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号