首页 | 本学科首页   官方微博 | 高级检索  
     


Double Cell Bulk Acoustic Wave Sensor for Ion Chromatographic Study of the Relation Between Magnesium and Growth of Organisms
Authors:You-tao Xie  Li-li Bao  Wan-zhi Wei
Affiliation:(1) Chemistry and Chemical Engineering College, Hunan University, Changsha, 410082, China , CN
Abstract:An ion chromatography (IC) method with a double cell bulk acoustic wave (DCBAW) detector was used for the detection of magnesium concentration decrease during the culture of whole cells. The detector parameters were optimized according to its theoretical equation, including the cell constants and the background conductance of the adjusting cell. A low temperature coefficient and thus high sensitivity were obtained by this method. The detection limit is 0.01 ppm, and the linear range is 0.05–50 ppm. A new mathematical model was developed to describe the growth of the cell. The baseline for this model is the concept that a noninteractive growth process occurs among perfectly substitutable nutrients if the locus of points of the substrate concentrations producing equal growth rate is linear. The monitoring results agreed well with the calculated results. Received: 17 July 1999 / Accepted: 23 August 1999
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号