Determination of the thermal and epithermal neutron sensitivities of an LBO chamber
Authors:Email authorEmail author  Hitoshi?Sato  Takuto?Shimazaki  Erika?Nakajima  Kei?Kotani  Mitsuru?Suda  Tsuyoshi?Hamano  Tsuyoshi?Kajimoto  Kenichi?Tanaka  Masaharu?Hoshi
Institution:1.Quantum Energy Applications, Graduate School of Engineering,Hiroshima University,Higashi-Hiroshima,Japan;2.Radiological Sciences,Ibaraki Prefectural University of Health Science,Inashiki-Gun,Japan;3.National Institute of Radiological Sciences,Chiba-Shi,Japan;4.Institute for Peace Science,Hiroshima University,Hiroshima,Japan;5.Delta Kogyo Co., Ltd.,Hiroshima,Japan
Abstract:An LBO (Li2B4O7) walled ionization chamber was designed to monitor the epithermal neutron fluence in boron neutron capture therapy clinical irradiation. The thermal and epithermal neutron sensitivities of the device were evaluated using accelerator neutrons from the 9Be(d, n) reaction at a deuteron energy of 4 MeV (4 MeV d-Be neutrons). The response of the chamber in terms of the electric charge induced in the LBO chamber was compared with the thermal and epithermal neutron fluences measured using the gold-foil activation method. The thermal and epithermal neutron sensitivities obtained were expressed in units of pC cm2, i.e., from the chamber response divided by neutron fluence (cm?2). The measured LBO chamber sensitivities were 2.23 × 10?7 ± 0.34 × 10?7 (pC cm2) for thermal neutrons and 2.00 × 10?5 ± 0.12 × 10?5 (pC cm2) for epithermal neutrons. This shows that the LBO chamber is sufficiently sensitive to epithermal neutrons to be useful for epithermal neutron monitoring in BNCT irradiation.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号