Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and discharges
Authors:Fani Zagkli  Panagiotis Chronopoulos  John Chiladakis
Institution:University Hospital of Patras, Department of Cardiology, Greece
Abstract:Ιn a 76-year old man with a dual-chamber ICD implanted five years ago, dizzy spells and significant bradycardia on Holter were not initially recognized as inhibition of bradycardia pacing, due to oversensing. Hospital admission was deemed necessary only after repetitive ICD shocks attributed to right ventricular pace-sense lead fracture. The need to ensure adequate ICD antibradycardia backup pacing in pacing-dependent patients when deleterious sensing errors occur, cannot be overemphasized.
Keywords:Bradycardia  ICD lead Fracture  Oversensing  Inappropriate shocks
本文献已被 ScienceDirect 等数据库收录!