首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Detection of Clostridium botulinum toxin A using a fiber optic-based biosensor.
Authors:R A Ogert  J E Brown  B R Singh  L C Shriver-Lake  F S Ligler
Institution:Center for Bio/Molecular Science and Engineering, Naval Research Laboratory, Washington, DC 20375.
Abstract:A rapid, sensitive, analytical method for the detection of Clostridium botulinum toxin has been developed. The fiber optic-based biosensor utilizes the evanescent wave of a tapered optical fiber for signal discrimination. A 50 mW argon-ion laser, which generates laser light at 514 nm, is used in conjunction with an optical fiber probe that is tapered at the distal end. Antibodies specific for C. botulinum are covalently attached to the surface of the tapered fiber. The principle of the system is a sandwich immunoassay using rhodamine-labeled polyclonal anti-toxin A immunoglobin G (IgG) antibodies for generation of the specific fluorescent signal. Various anti-toxin antibodies were immobilized to the fibers. Affinity-purified polyclonal horse anti-toxin A antibodies performed better than the IgG fraction from the same horse serum or than the monoclonal anti-toxin A antibody BA11-3. Botulinum toxin could be detected within a minute, at concentrations as low as 5 ng/ml. The reaction was highly specific and no response was observed against tetanus toxin.