首页 | 本学科首页   官方微博 | 高级检索  
     


Constraints and fixation for implanted joint replacements
Affiliation:1. Key Laboratory of Bionic Engineering, Ministry of Education, Jilin University, Changchun, China;2. Institute of Orthopaedic & Musculoskeletal Science, University College London, Royal National Orthopaedic Hospital, Stanmore HA7 4LP, UK;3. State Key Laboratory for Manufacturing System Engineering, School of Mechanical Engineering, Xi’an Jiaotong University, Xi’an, China;4. State Key Laboratory of Metal Matrix Composites, Shanghai Jiao Tong University, Shanghai, China
Abstract:Observations made on both cemented and uncemented joint prostheses after about two years of use show a layer of fibrous tissue next to the bone. This fibrous layer smooths over surface features up to l mm in size, and must be assumed to be weak in tension and shear. Prosthesis-bone interfaces should therefore be designed to transmit all forces acting on the prosthesis as compressive stresses. The forces acting across prostheses are controlled by the constraints exerted by the articulating surfaces; if the useful ligaments present are allowed to do their job of transmitting tension, the articulating surfaces can be designed to transmit only such forces and moments as can safely be transmitted across the prosthesis-bone interfaces.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号