High-performance liquid chromatography coupled with electrospray tandem mass spectrometry (LC/MS/MS) method for the simultaneous determination of diazepam, atropine and pralidoxime in human plasma |
| |
Authors: | Abbara Chadi Bardot Isabelle Cailleux Annie Lallement Guy Le Bouil Anne Turcant Alain Clair Pascal Diquet Bertrand |
| |
Affiliation: | Université d'Angers, UFR médecine, Angers cedex F-49045, France. chadi.abbara@gmail.com |
| |
Abstract: | A high-performance liquid chromatography coupled with electrospray tandem mass spectrometry (LC/MS/MS) procedure for the simultaneous determination of diazepam from avizafone, atropine and pralidoxime in human plasma is described. Sample pretreatment consisted of protein precipitation from 100microl of plasma using acetonitrile containing the internal standard (diazepam D5). Chromatographic separation was performed on a X-Terra MS C8 column (100mmx2.1mm, i.d. 3.5microm), with a quick stepwise gradient using a formate buffer (pH 3, 2mM) and acetonitrile at a flow rate of 0.2ml/min. The triple quadrupole mass spectrometer was operated in positive ion mode and multiple reaction monitoring was used for drug quantification. The method was validated over the concentration ranges of 1-500ng/ml for diazepam, 0.25-50ng/ml for atropine and 5-1000ng/ml for pralidoxime. The coefficients of variation were always <15% for both intra-day and inter-day precision for each analyte. Mean accuracies were also within +/-15%. This method has been successfully applied to a pharmacokinetic study of the three compounds after intramuscular injection of an avizafone-atropine-pralidoxime combination, in healthy subjects. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|