The Preparation of Silica Gel Thin Layers with a New Apparatus
Authors:D. T. Organisciak  J. D. Klingman  W. A. Tanski
Affiliation:1. Department of Biochemistry , Wright State University School of Medicine , Dayton, Ohio, 45431;2. Department of Biochemistry , State University of New York at Buffalo , Buffalo, New York, 14214;3. Department of Technical Services , State University of New York at Buffalo , Buffalo, New York, 14214
Abstract:An apparatus for the preparation of uniform thin layer chromatography gel layers is described. The apparatus employs rubber cushion runners to compensate for differences in plate thickness and a gel applicator which functions independently of plate edge variability. Consistently uniform layers are prepared by careful establishment of the plate-to-applicator distance. Silica gel layers averaged 87% of the applied thickness with a variability of ±2% within a single run and ±8% between independent runs.