Simultaneous determination of ABT-888, a poly (ADP-ribose) polymerase inhibitor,and its metabolite in human plasma by liquid chromatography/tandem mass spectrometry
Authors:Richard Wiegand  Jianmei Wu  Xianyi Sha  Patricia LoRusso  Jing Li
Institution:Barbara Ann Karmanos Cancer Institute, Wayne State University, Detroit, MI 48201, USA
Abstract:A reversed-phase liquid chromatography coupled to tandem mass spectrometry (LC–MS/MS) method was developed and validated for simultaneous determination of ABT-888 and its major metabolite (M8) in human plasma. Sample preparation involved a liquid–liquid extraction by the addition of 0.25 ml of plasma with 10 μl of 1 M NaOH and 1.0 ml ethyl acetate containing 50 ng/ml of the internal standard zileuton. The analytes were separated on a Waters XBridge C18 column using a gradient mobile phase consisting of methanol/water containing 0.45% formic acid at the flow rate of 0.2 ml/min. The analytes were monitored by tandem mass spectrometry with electrospray positive ionization. Linear calibration curves were generated over the ABT-888 and M8 concentration ranges of 1–2000 ng/ml in human plasma. The lower limits of quantitation (LLOQ) were 1 ng/ml for both ABT-888 and M8 in human plasma. The accuracy and within- and between-day precisions were within the generally accepted criteria for bioanalytical method (<15%). This method was successfully employed to characterize the plasma concentration–time profile of ABT-888 after its oral administration in cancer patients.