首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of an electrochemiluminescent homogenous immunoassay and an ELISA for monitoring productivity in mammalian cell bioreactors
Authors:Per I. Lindell  Lee Hall  Tai-Guang Wu  Charles L. Cooney
Affiliation:(1) Biotechnology Process Engineering Center, Massachusetts Institute of Technology, 02139 Cambridge, MA, USA;(2) IGEN Inc., 1530 East Jefferson St., 20852 Rockville, MD, USA
Abstract:Summary An Enzyme Linked Immuno Sorbent Assays (ELISA) and an Electro-chemiluminescent Immuno Assay (CIA) are compared for the purpose of monitoring product formation in mammalian cell bioreactors. The ELISA had a relative standard deviation of 10%, compared to 8% for the CIA . The CIA was found to be a fast and accurate alternative to the ELISA. The use of more than one immuno assay format was also shown to provide additional insight into process performance.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号