首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Multiregional evaluation of the SimPlate heterotrophic plate count method compared to the standard plate count agar pour plate method in water
Authors:Jackson R W  Osborne K  Barnes G  Jolliff C  Zamani D  Roll B  Stillings A  Herzog D  Cannon S  Loveland S
Institution:Cobb County-Marietta Water Authority, Acworth, Georgia 30101, USA.
Abstract:A new SimPlate heterotrophic plate count (HPC) method (IDEXX Laboratories, Westbrook, Maine) was compared with the pour plate method at 35 degrees C for 48 h. Six laboratories tested a total of 632 water samples. The SimPlate HPC method was found to be equivalent to the pour plate method by regression analysis (r = 0. 95; y = 0.99X + 0.06).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏