Validation of a dual energy X-ray absorptiometer: measurement of bone mass and soft tissue composition |
| |
Authors: | Ole Rintek Madsen Jens-Erik Beck Jensen and Ole Helmer S?rensen |
| |
Institution: | (1) Osteoporosis Research Centre, Copenhagen Municipal Hospital, University of Copenhagen, ?ster Farimagsgade 5, DK-1399 Copenhagen K, Denmark., DK |
| |
Abstract: | We investigated the reproducibility of total and regional body composition measurements performed on a dual energy X-ray
absorptiometer (DXA). A group of 38 women aged 21–81 (mean 52. 4) years was scanned twice with repositioning to determine
intra-observer reproducibility of measurements of bone mineral density (BMD, g · cm−2), bone mineral content (BMC, g), lean mass (LM, kg) and fat mass (FM, kg) of the total body and of the major subregions of
the body. In addition, the ability of the DXA machine to detect changes in LM and FM (simulated by placing 11.1 and 22.3 kg
porcine lard on the body of 11 subjects) was examined. Coefficients of variations calculated from the root mean square averages
of individual standard deviations were as follows (BMD, BMC, FM, LM): 1.4%, 1.1%, 1.4%, 1.7% (total body), 2.2%, 2.1%,-,-
(head), 2.8%, 2.8%, 2.0%, 2.2% (trunk), 3.6%, 3.9%, 4.0%, 4.9% (arms), 2.7%, 1.3%, 2.6%, 2.8% (legs). Percentage fat (%fat)
of exogenous lard was 81.3 (SD 3.5)% as assessed by the absorptiometer which corresponded well with the result of chemical
analysis (82.8%). Estimated %fat of exogenous lard was not influenced by initial body mass or percentage body fat. Percentages
of expected mean values with 11.1 kg lard placed on the body were 99.9 (SD 0.3) for body mass, 100.5 (SD 2.1) for LM, and
99.5 (SD 3.5) for FM. BMD was overestimated by 3.2% (P < 0.005) with 11.1 kg lard on the body. BMD as well as BMC increased significantly with 22.3␣kg lard on the body (P < 0.005). The results showed that BMD, BMC, LM, and FM of the total body were precisely estimated by the DXA machine used.
Regional measurements were less precise. Changes in total body soft tissue composition were precisely and accurately estimated.
The lard placed on the body falsely affected BMD and BMC measurements. Changes in body mass could have a similar effect.
Accepted: 6 January 1997 |
| |
Keywords: | Bone mineral density Dual energy X-ray absorptiometry Fat mass Lean mass Reproducibility |