首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Development of a flow‐injection analysis system with fluorescence detection for gatifloxacin determination in organized medium
Authors:Monica F Lima Vaz  João Vitor F de Oliveira  Ricardo J Cassella  Wagner F Pacheco
Institution:Departamento de Química Analítica, Universidade Federal Fluminense, Niterói, RJ, Brazil
Abstract:This work reports the development and optimization of a flow injection analysis system with fluorescence detection (FIA–FLUO) for gatifloxacin (GFX) determination in organized medium. The analytical system was based on the enhanced fluorescence of gatifloxacin in micellar medium containing sodium dodecyl sulfate (SDS) at pH 6.0. The influence of physical (carrier flow rate, sample volume and volume of reaction coil) and chemical (pH, concentration of buffer and concentration of SDS) parameters that could affect the performance of the FIA system was evaluated in order to reach optimum conditions in terms of sensitivity and analytical throughput. Under optimized conditions, the FIA–FLUO system allowed the injection of 40 samples per hour with a limit of quantification of 72 µg/L and a RSD of 3.5% at 0.20 mg/L. Real samples of commercial pharmaceutical formulations containing GFX were analyzed, and no statistical difference was observed between the results obtained using the developed system and those obtained using the reference method based on high‐performance liquid chromatography with UV detection. Copyright © 2014 John Wiley & Sons, Ltd.
Keywords:flow injection analysis  fluorescence  gatifloxacin  pharmaceutical formulations
设为首页 | 免责声明 | 关于勤云 | 加入收藏