首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Imaging of chemiluminescent signals with cooled CCD camera systems
Authors:Chris S Martin  Irena Bronstein
Abstract:We investigated imaging of chemiluminescent signals from 1,2-dioxetanes with cooled CCD cameras. Non-radioactive detection methods for biomolecules utilizing these chemiluminescent substrates for alkaline phosphatase have been developed. Applications which have been successfully adapted to this technology include Southern and Northern blotting, immunoblotting, ELISA methods and DNA sequencing. Dephosphorylation of the dioxetane CSPD by alkaline phosphatase generates an unstable anion that decomposes resulting in light production. The wavelength of the emitted light is approximately 460nm. We have utilized Photometrics Star and MXC 200L cooled CCD cameras for direct imaging of chemiluminescent signals. Benefits of utilizing a CCD detector include rapid data digitization and more accurate quantitation of chemiluminescent signals compred to film-based densitometry owing to the significantly greater dynamic range. Chemiluminescent images from dot blots of biotinylated DNA, Southern blots and DNA sequencing gel blots were obtained. In a chemiluminescent microtitre plate assay, serial dilutions of alkaline phosphatase spanning four orders of magnitude can be detected. Our results indicate that the digitization of chemiluminescent signal data with cooled CCD cameras is an excellent alternative to 32P detection methods utilizing storage phosphor screen imaging systems.
Keywords:Chemiluminescence  CCD camera  imaging  1  2-dioxetanes
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号