首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of a new semi-automated Hydragel 11 von Willebrand factor multimers assay kit for routine use
Authors:Marika Pikta,Timea Szanto,Margus Viigimaa,Sandra Lejniece,D&#x  rta Balode,Kadri Saks,Valdas Banys
Abstract:BackgroundAccurate diagnosis and classification of von Willebrand disease (VWD) are essential for optimal management. The von Willebrand factor multimers analysis (VWF:MM) is an integral part of the diagnostic process in the phenotypic classification, especially in discrepant cases. The aim of this study was to evaluate the performance of a new Hydragel 11VWF multimer assay (H11VW).MethodsAnalytical performance characteristics such as repeatability (intra-assay variability, in gel between track variation), reproducibility (inter-assay variability, between gel variation), sensitivity, EQA performance and differences between two commercially available VWF:MM kits (H5VW and H11VW) were analysed in healthy volunteers'' plasmas using in-house prepared reference plasma.ResultsRepeatability and reproducibility results of H11VW demonstrated acceptable and equivalent performance with previously verified H5VW. Participation in EQA was successful. No statistically significant difference was detected between H5VW and H11VW kits for different fractions of multimers: LMWM p=0.807; IMWM p=0.183; HMWM p=0.774.ConclusionsH11VW demonstrated acceptable analytical performance characteristics. H11VW kit conveniently offers a more significant number of samples on a single gel. H5VW and H11VW kits can be used in daily practice interchangeably.
Keywords:electrophoresis   multimers   von Willebrand factor   von Willebrand disease
设为首页 | 免责声明 | 关于勤云 | 加入收藏