Measurement of serum gentamycin levels by rapid tests
Authors:S M Hussain Qadri  R L Gillum  Debbie Boyd
Institution:(1) University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma, USA;(2) Oklahoma Memorial Hospital and Clinics, Oklahoma City, Oklahoma, USA;(3) Department of Pathology, The University of Oklahoma Health Sciences Center, P.O. Box 26901, 73190 Oklahoma City, OK, USA
Abstract:Gentamicin is a very useful antimicrobial agent for the treatment of serious infections caused by gram-negative bacteria. However, it's low therapeutic index and potential ototoxic and nephrotoxic side effects necessitate frequent determinations of serum concentration to assist in maintaining therapeutic levels and avoiding toxic levels. Two bioassays and a latex agglutination inhibition card (LAIC) test were evaluated to determine gentamicin levels in nearly 100 patient sera. Results were compared with a radioimmunoassay (RIA). Two bioassays, the Bio-Monitor and the GentaSak, gave correlation coefficients of 0.987 and 0.982, respectively. The correlation coefficient for the LAIC test was 0.987. All three tests compared well with RIA in accurately detecting gentamicin levels in patient as well as simulated sera. The LAIC test, however, was more rapid, giving results within half an hour whereas bioassays required 6–8 hours for completion. The LAIC test was also found to be more economical. It provides a suitable alternative to RIA procedures in small laboratories and for performing lsquostatrsquo tests since batching is not necessary.
Keywords:
本文献已被 SpringerLink 等数据库收录!