An Improved Method with High Anti-interference Ability for R Peak Detection in Wearable Devices
Affiliation:1. Siyami Ersek Thoracic and Cardiovascular Surgery Training and Research Hospital, Department of Cardiology, Istanbul, Turkey;2. Gulhane Military Medical Academy, Haydarpasa Training Hospital Department of Cardiology, Istanbul, Turkey
Abstract:The rapid development of the wearable electrocardiogram monitoring equipment increases the requirements for R peak detection in wearable devices. An improved method called ISC algorithm is proposed with high anti-interference ability for R peak detection in wearable devices based on a simple basic algorithm called SC algorithm. The proposed method is characterized by using the updated amplitude selection threshold, updated slope comparison threshold and RR interval judgement to reduce false positives and false negatives. For data from MIT-BIH Arrhythmia Database, the positive predictivity P+ of ISC algorithm can reach 99.12%, and the sensitivity Se of ISC algorithm is more than 95%. For MIT-BIH Noise Stress Test Database, the accuracy of ISC algorithm for both sensitivity Se and positive predictivity P+ can exceed 94% under three common noise, baseline wander, muscle artifact, and electrode motion artifact, where the positive predictivity P+ of ISC algorithm is 44.46% higher than that of SC algorithm on average. For wearable devices in exercise, even under the exercise intensity of 7 km per hour, the average positive predictivity P+ of ISC algorithm is 99.32%, which is 60.93% higher than that of SC algorithm. The high anti-interference ability shows that ISC algorithm is suitable for R peak detection in wearable devices.
Keywords:High anti-interference ability  R peak detection  RR interval judgement  Updated amplitude selection threshold  Updated slope comparison threshold  Wearable device
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号