首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Energy and dose dependence of GafChromic EBT3-V3 film across a wide energy range
Institution:St. Luke’s Radiation Oncology Network (SRLON), Highfield Road, Rathgar, Dublin, Ireland
Abstract:AimTo determine the energy and dose dependence of GafChromic EBT3-V3 film over an energy range 0.2 mm Al HVL to 6 MV.BackgroundThe decay scheme of a brachytherapy source may be complex and the spectrum of energy can be wide. LiF TLDs are the golden standard recommended for dosimetric measures in brachytherapy, for their energy independence, but TLDs could be not available in some centres. An alternative way to perform dose measurements is to use GafChromic films, but they show energy dependence.Methods and materialsFilms have been irradiated at increasing dose with three different beams: 6 MV beam, TPR20, 10 = (0.684 ± 0.01), HVL = (2.00 ± 0.01)mmAl and HVL = (0.20 ± 0.01)mmAl. Calibration curves were generated using the same dose range (0cGy to 850cGy) for the three energies. Using the 6 MV calibration curve as reference, the film response in terms of net optical density (OD) was evaluated.ResultsThe difference in the calibration curve obtained by irradiating the film with 6 MV and 2 mm Al HVL energy beams is less than 3 %, within the calibration uncertainty, in the dose range 500-850cGy. The OD of EBT3-V3 film is significantly lower at 0.2 mmAl HVL compared to 6 MV, showing differences up to 25 %.ConclusionWithin the range 6 MV-2 mm Al HVL and dose higher than 500cGy, GafChromic EBT3-V3 films are energy independent. In this dose range, films can be calibrated in a simple geometry, using a 6 MV Linac beam, and can be used for brachytherapy sources dose measures. The use of EBT3 films can be extended to reference dosimetry in Ir-192 clinical brachytherapy.
Keywords:GafChromic EBT3-V3  Dosimetry  Energy dependence  Dose dependence  IPEM standard  DIN6809-4 standard
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号