首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Technique validation by liquid chromatography for the determination of acyclovir in plasma
Authors:Fernández Marcos  Sepúlveda Jacqueline  Aránguiz Teobaldo  von Plessing Carlos
Institution:Pharmacy Department, Pharmacy Faculty, Universidad de Concepción, Casilla 237, Concepción, Chile.
Abstract:In this research project, a high-performance liquid chromatography (HPLC) method was developed for the determination of acyclovir (ACV) in plasma. The plasma samples, recharged with acyclovir and in presence of 5'-N-methylcarboxyamidoadenosine (MECA) as an internal standard, were purified using a solid-phase extraction technique with Waters Oasis HLB columns. The separation of the components from the extract was carried out in a LiChrospher 100 RP-18 column for further ultraviolet detection at a wavelength range of 250-260 nm. The mobile phase composition was 18% acetonitrile, sodium dodecylsulphate 5 mM and phosphate buffer at pH 2.6 with an analysis time of 13 min per sample. The average retention time for acyclovir was of 5.0 min and for the internal standard 11.2 min. The calibration curve was linear ranging between 0.05 and 1.80 microg/ml. The detection limit was 0.006 microg/ml with a quantification limit of 0.020 microg/ml. The ACV recuperation percentage for 250 microl of plasma was between 94.7 and 109.7% with a coefficient of variation not higher than 5.2%. This method was developed and validated for use in bioavailability and bioequivalence studies.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏