Determination of ranitidine in urine by capillary electrophoresis-electrochemiluminescent detection |
| |
Authors: | Gao Ying Tian Yiling Sun Xiuhua Yin Xue-Bo Xiang Qian Ma Ge Wang Erkang |
| |
Institution: | Department of Chemistry, Tianjin University, Tianjin, China. |
| |
Abstract: | The fast analysis of ranitidine is of clinical importance in understanding its efficiency and a patient's treatment history. In this paper, a novel determination method for ranitidine based on capillary electrophoresis-electrochemiluminescence detection is described. The conditions affecting separation and detection were investigated in detail. End-column detection of ranitidine in 5 mM Ru(bpy)(3)(2+) solution at applied voltage of 1.20 V was performed. Favorable ECL intensity with higher column efficiency was achieved by electrokinetic injection for 10s at 10 kV. The R.S.D. values of ECL intensity and migration time were 6.38 and 1.84% for 10(-4) M and 6.01 and 0.60% for 10(-5) M, respectively. A detection limit of 7 x 10(-8) M (S/N=3) was achieved. The proposed method was applied satisfactorily to the determination of ranitidine in urine in 6 min. |
| |