首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Biomechanical evaluation of porous biodegradable scaffolds for revision knee arthroplasty
Authors:Alexandre Terrier  Marjan Sedighi-Gilani  Alireza Roshan Ghias  Line Aschwanden  Dominique P Pioletti
Institution:1. Laboratory of Biomechanical Orthopedics , Ecole Polytechnique Federal de Lausanne , Lausanne, Switzerland alexandre.terrier@epfl.ch;3. Laboratory of Biomechanical Orthopedics , Ecole Polytechnique Federal de Lausanne , Lausanne, Switzerland
Abstract:Tibial bone defect is a critical problem for revision knee arthroplasty. Instead of using metallic spacer or cement, biodegradable scaffolds could be an alternative solution. A numerical model of a revision knee arthroplasty was thus developed to estimate the mechanical resistance of the scaffold in this demanding situation. The tibia, scaffold, and prosthesis were represented by simplified parameterised geometries. The maximal gait cycle force was applied asymmetrically to simulate a critical loading. Several parameters were analysed: 1) inter-individual variability, 2) cortical bone stiffness, 3) cortical bone thickness, 4) prosthesis fixation quality, and 5) scaffold thickness. The calculated scaffold strain was compared to its experimental ultimate strain. Among the tested parameters, failure was only predicted with scaffold thickness below 5 mm. This study suggests that biodegradable bone scaffolds could be used to fill bone defects in revision knee arthroplasty, but scaffold size seems to be the limiting factor.
Keywords:bone scaffold  prosthesis  knee  revision arthroplasty
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号