首页 | 本学科首页   官方微博 | 高级检索  
   检索      


An amperometric bi-enzyme sensor for determination of formate using cofactor regeneration
Authors:Mak Karen K W  Wollenberger Ulla  Scheller Frieder W  Renneberg Reinhard
Institution:Sino-German Nano-Analytical Lab (SiGNAL) and Department of Chemistry, The Hong Kong University of Science and Technology, Clear Water Bay, Kowloon, Hong Kong. karen@ust.hk
Abstract:A biosensor for detection of formate at submicromolar concentrations has been developed by co-immobilizing formate dehydrogenase (FDH, E.C. 1.2.1.2), salicylate hydroxylase (SHL, E.C. 1.14.13.1) and NAD(+) linked to polyethylene glycol (PEG-NAD(+)) in a poly(vinyl alcohol) (PVA) matrix in front of a Clark-electrode. The principle of the bi-enzyme scheme is as follows: formate dehydrogenase converts formate into carbon dioxide using PEG-NAD(+). Corresponding PEG-NADH produced is then oxidized to PEG-NAD(+) by salicylate hydroxylase using sodium salicylate and oxygen. The oxygen consumption is monitored with the Clark-electrode. The advantages of this biosensor approach are the effective re-oxidation of PEG-NADH, and the entrapment of PEG-NAD(+) resulting in avoiding the addition of expensive cofactor to the working medium for each measurement. This bi-enzyme sensor has achieved a linear range of 1-300 microM and a detection limit of 1.98 x 10(-7) M for formate (S/N=3), with the response time of 4 min. The working stability is limited to 7 days due to the inactivation of the enzymes. Only sodium salicylate was needed in milli-molar amounts.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏