首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Interference elimination of an amperometric glucose biosensor using poly(hydroxyethyl methacrylate) membrane
Authors:Azila Abdul‐Aziz  Fui‐Ling Wong
Institution:Department of Bioprocess Engineering, Universiti Teknologi Malaysia, Skudai, Johor, Malaysia
Abstract:A permselective membrane fabricated from photo‐cross‐linked poly(hydroxyethyl methacrylate) (pHEMA) was studied as a potential selective membrane that can eliminate electrochemical interferences commonly faced by a hydrogen peroxide‐based biosensor. The quantitative selection of the permselective membrane was based on the permeabilities of hydrogen peroxide and acetaminophen (AC). AC was used as a model of the interfering substance due to its neutral nature. pHEMA membrane with the cross‐linking ratio of 0.043 was found to achieve a selectivity of hydrogen peroxide over AC of 10, while maintaining an acceptable degree of hydrogen peroxide response. A two‐layer glucose biosensor model consisting of glucose oxidase entrapped within a freeze‐thawed poly(vinyl alcohol) matrix and the cross‐linked pHEMA membrane was challenged with AC, ascorbic acid and uric acid. 0.2 mM AC and 0.2 mM ascorbic acid were completely eliminated. However, 0.2 mM uric acid could not be completely eliminated and still gave a bias of approximately 6.6% relative to 5 mM glucose. The results showed that cross‐linked pHEMA was quite promising as an interference eliminating inner membrane.
Keywords:Electrochemical interference  Hydrogen peroxide‐based biosensor  Poly(hydroxyethyl methacrylate)
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号