首页 | 本学科首页   官方微博 | 高级检索  
     


Development of a sandwich format, amperometric screen-printed uric acid biosensor for urine analysis
Authors:Kanyong P  Pemberton R M  Jackson S K  Hart J P
Affiliation:Centre for Research in Biosciences, University of the West of England, Frenchay Campus, Bristol BS16 1QY, UK.
Abstract:A screen-printed carbon electrode (SPCE) incorporating the electrocatalyst cobalt phthalocyanine (CoPC), fabricated using a water-based ink formulation, has been investigated as the base transducer for a uric acid biosensor. A sandwich biosensor was fabricated by first depositing cellulose acetate (CA) onto this transducer (CoPC-SPCE), followed by uricase (UOX) and finally a polycarbonate (PC) membrane; this device is designated PC-UOX-CA-CoPC-SPCE. This biosensor was used in conjunction with chronoamperometry to optimize the conditions for the analysis of urine: temperature, 35°C; buffer, pH 9.2; ionic strength, 50 mM; uricase, 0.6 U; incubation time, 180 s. The proposed biosensor was applied to urine from a healthy subject. The precision determined on unspiked urine (n=6) was 5.82%. Urine was fortified with 0.225 mM UA, and the resulting precision and recovery were 4.21 and 97.3%, respectively. The linear working range of the biosensor was found to be 0.015 to 0.25 mM (the former represents the detection limit), and the sensitivity was calculated to be 2.10 μA/mM.
Keywords: