首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of oxytetracycline levels in rainbow trout serum on a biphenyl column using high-performance liquid chromatography
Authors:Miller Ron A  Reimschuessel Renate  Carson Mary C
Affiliation:U.S. Food and Drug Administration, Center for Veterinary Medicine, Office of Research, 8401 Muirkirk Road, Laurel, MD 20708, USA. Ron.Miller@fda.hhs.gov
Abstract:We developed a simple and sensitive high-performance liquid chromatography method on a biphenyl column to determine oxytetracycline (OTC) levels in rainbow trout serum. The assay used deproteination, filtration, and subsequent separation on a reverse-phase biphenyl column, with UV detection at 355 nm. OTC (7.8-7.9 min) was completely resolved from structurally similar riboflavin (10.4-10.5 min), a common feed supplement. Estimated limits of detection and quantitation of OTC were 0.01 and 0.04 microg/mL, respectively. The average recovery for OTC was 102% with a R.S.D. of 8.34%. Calibration standards were linear from 0.01 to 10 microg/mL.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号