首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of iohexol clearance by high-performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS)
Authors:Lee Soo-Youn  Chun Mi-Ryung  Kim Dae-Joong  Kim Jong Won
Affiliation:Department of Laboratory Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, South Korea.
Abstract:We have developed a simple, rapid, and accurate HPLC-MS/MS method for the determination of iohexol in serum. The column used was a Zorbax Eclipse XDB-C8 (100 mm x 2.1 mm i.d., 3.5 microm). Mobile phases consisted of water containing 2mM ammonium acetate and 0.1% formic acid (A) and methanol containing 2 mM ammonium acetate and 0.1% formic acid (B). After simple protein precipitation with ZnSO4, serum samples were mixed with I.S. (bromperidol) and centrifuged for 3 min. The obtained extraction recovery at three levels was 94.6-107.4%. Quantitative analysis was performed in the multiple reaction-monitoring mode (m/z 822.0-->804.0 for iohexol, 420.1-->122.7 for I.S.) with the total running time of 3 min for each sample. The assay was linear between 0.5 and 1500 microg/mL (r2 > 0.997). The intra- and inter-assay coefficient of variations were 2.4-6.2% and 5.5-6.5%, respectively. Our method provided sufficient analytical range and specificity for the 210 clinical samples analyzed.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号