首页 | 本学科首页   官方微博 | 高级检索  
     


Monte Carlo simulation for Neptun 10 PC medical linear accelerator and calculations of output factor for electron beam
Authors:Mohammad Taghi Bahreyni Toossi  Mehdi Momennezhad  Seyed Mohammad Hashemi
Affiliation:1. Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran;2. Nuclear Medicine Research Center, Imam Reza Hospital, Mashhad University of Medical Sciences, Mashhad, Iran
Abstract:AimExact knowledge of dosimetric parameters is an essential pre-requisite of an effective treatment in radiotherapy. In order to fulfill this consideration, different techniques have been used, one of which is Monte Carlo simulation.Materials and methodsThis study used the MCNP-4Cb to simulate electron beams from Neptun 10 PC medical linear accelerator. Output factors for 6, 8 and 10 MeV electrons applied to eleven different conventional fields were both measured and calculated.ResultsThe measurements were carried out by a Wellhofler-Scanditronix dose scanning system. Our findings revealed that output factors acquired by MCNP-4C simulation and the corresponding values obtained by direct measurements are in a very good agreement.ConclusionIn general, very good consistency of simulated and measured results is a good proof that the goal of this work has been accomplished.
Keywords:Simulation   Monte Carlo   Linear accelerator   Dosimetry