The effect of scatter correction and radius of rotation on semiquantitative measurements in SPECT 123I-FP-CIT imaging. A phantom study
Institution:1. Division of Advanced Cardiac Imaging, King Saud bin Abdulaziz University for Health Sciences, Riyadh, Saudi Arabia;2. Section of Nuclear Medicine and Diagnostic Imaging, Division of Human Health, International Atomic Energy Agency, Vienna, Austria;3. Columbia University Medical Center, New York-Presbyterian Hospital, New York, NY, United States;4. Department of Medicine, McMaster University, Hamilton, Ontario, Canada;5. Quanta Diagnóstico & Terapia, Curitiba, Brazil;6. Department of Cardiology, All India Institute of Medical Sciences, New Delhi, India;7. Department of Nuclear Medicine, Royal Melbourne Hospital, University of Melbourne, Melbourne, Australia;8. Department of Radiology, Columbia University Medical Center, New York-Presbyterian Hospital, NY, New York, United States
Abstract:PurposeThe high energy emissions of 123I and the suboptimal radius of rotation affect the semiquantitative measurements performed during 123I-FP-CIT tomographic imaging. An in-house extra low cost striatum phantom with brain and striatum compartments was constructed and was used to study the effects of Triple Energy Window scatter correction (TEW-SC) and radius of rotation on the Specific Binding Ratio (SBR) measurements.Materials and methodsThe phantom compartments were filled with radioactive 123I solutions with varying concentrations, in a series of experiments. Tomographic images were acquired at six different radii of rotation, with and without TEW-SC and the SBRs were calculated using appropriate regions of interest, as in clinical imaging.ResultsSBRs decreased with increasing radius of rotation in both non-SC and TEW-SC images, the decrease being more pronounced in the latter. The application of TEW-SC increases SBR values by 40% on average. A maximum %Recovery of 42.7% of the true SBR value was achieved in the non-SC images, which increased to 64.6% after TEW-SC. Appropriate correction factors (CF) were calculated in order to make the SBR values independent on the radius of rotation, which could be used to correct SBR values obtained from tomographic acquisitions with suboptimal radius of rotation.ConclusionThe use of appropriate CF can provide more consistent SBR values and a more meaningful comparison between SBRs calculated from images acquired at different radii of rotation.
Keywords:Specific Binding Ratio  Radius of rotation  Scatter correction
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号