首页 | 本学科首页   官方微博 | 高级检索  
     


Feasibility of using a commercial collapsed cone dose engine for 1.5T MR-LINAC online independent dose verification
Affiliation:1. Department of Radiation Oncology, Sun Yat-Sen University Cancer Center, State Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Guangzhou 510060, China;2. School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China
Abstract:PurposeTo validate the feasibility and accuracy of commonly used collapsed cone (CC) dose engine for Elekta Unity 1.5T MR-LINAC online independent dose verification.Materials and MethodsThe Unity beam model was built and commissioned in RayStation treatment planning system with CC dose engine. Four AAPM TG-119 test plans were created and measured with ArcCHECK phantom for comparison, another thirty patient plans from six tumor sites were also included. The dosimetric criteria for various ROIs and 3D gamma passing rates were quantitatively evaluated, and the effects of magnetic field and dose deposition type on the dose difference between two systems were further analyzed.ResultsArcCHECK based measurement showed a clear magnetic field induced profile shift between CC with both measurement and GPUMCD. For clinical plans, gamma passing rates with criteria (3%, 3 mm) between GPUMCD and CC large than 90% can be achieved for most tumor sites except esophagus and lung cases, the mean dose difference of 3% can be satisfied for most ROIs from all tumor sites. The magnetic field caused a large dose impact on low density areas, the average gamma passing rates were improved from 85.54% to 96.43% and 87.40% to 99.54% for esophagus and lung cases when the magnetic field effect was excluded.ConclusionsIt is feasible to use CC dose engine as a secondary dose calculation tool for Elekta Unity system for most tumor sites, while the accuracy is limited and should be used carefully for low density areas, such as esophagus and lung cases.
Keywords:MR-LINAC  Secondary independent dose verification  Collapsed cone dose calculation  GPUMCD
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号