End-to-end dosimetric audit: A novel procedure developed for Irish HDR brachytherapy centres
Institution:1. Azienda Sanitaria Universitaria Friuli Centrale, Udine, Italy;2. Galway University Hospitals, Medical Physics and Bioengineering, Galway, Ireland;3. Health Service Executive, Department of Medical Physics and Bioeengineering, Galway, Ireland
Abstract:PurposeA dosimetric audit of Ir-192 high dose rate (HDR) brachytherapy remote after-loading units was carried out in 2019. All six brachytherapy departments on the island of Ireland participated in an end-to-end test and in a review of local HDR dosimetry procedures.Materials and methodsA 3D-printed customised phantom was created to position the following detectors at known distances from the HDR source: a Farmer ionization chamber, GafChromic film and thermoluminescent dosimeters (TLDs). Dedicated HDR applicator needles were used to position an Ir-192 source at 2 cm distance from these detectors. The end-to-end dosimetry audit pathway was performed at each host site and included the stages of imaging, applicator reconstruction, treatment planning and delivery. Deviations between planned and measured dose distributions were quantified using gamma analysis methods. Local procedures were also discussed between auditors and hosts.ResultsThe mean difference between Reference Air Kerma Rate (RAKR) measured during the audit and RAKR specified by the vendor source certificate was 1.3%. The results of end-to-end tests showed a mean difference between calculated and measured dose of 2.5% with TLDs and less than 0.5% with Farmer chamber measurements. GafChromic films showed a mean gamma passing rates of >95% for plastic and metal applicators with 2%/1 mm global tolerance criteria.ConclusionsThe results of this audit indicate dosimetric consistency between centres. The ‘end to end’ dosimetry audit methodology for HDR brachytherapy has been successfully implemented in a multicentre environment, which included different models of Ir-192 sources and different treatment planning systems.The ability to create a 3D-printed water-equivalent phantom customised to accurately position all three detector types simultaneously at controlled distances from the Ir-192 source under evaluation gives good reproducibility for end-to-end methodology.
Keywords:Dosimetric audit  HDR brachytherapy  3D printable materials  Ir-192 dosimetry  TLD  GafChromic films  Farmer chamber
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号