A sensitive, specific, accurate and reproducible LC–MS–MS method was developed and validated for the simultaneous quantitation of the prodrug fosinopril and its active drug fosinoprilat in human serum. The method employed acidification of the serum samples to minimize the hydrolysis of fosinopril to fosinoprilat prior to purification by solid-phase extraction to isolate the two analytes and the two internal standards from human serum. The extracted samples were analyzed by turbo ionspray LC–MS–MS in the positive ion mode. Chromatography was performed on a polymer-based C18 column (Asahipak™ ODP PVA-C18, 2×50 mm) using gradient elution with methanol and 10 mM ammonium acetate, pH 5.5. The calibration curve, 1.17 to 300 ng/ml, was fitted to a weighted (l/x) linear regression model. Serum quality control (QC) samples used to gauge the accuracy and precision of the method were prepared at concentrations of 5.00, 100, 250 and 500 ng/ml of each analyte. The inter-assay accuracies were within 6% (DEV) for both analytes. The intra- and inter-assay precisions were within 7% and 11% (RSD), respectively, for both analytes. The hydrolysis of fosinopril to fosinoprilat during sample processing was ≤6%. This degree of conversion would cause little error in the analysis of post-dose serum samples since such samples are known to contain low levels of the prodrug compared to the drug.  相似文献   