1.

Objective

The purpose of this study was to determine if anodal stimulation accounts for failure to benefit from cardiac resynchronization therapy (CRT) in some patients.

Background

Approximately 30-40% of patients with moderate to severe heart failure do not have symptomatic nor echocardiographic improvement in cardiac function following CRT. Modern CRT devices allow the option of programming left ventricular (LV) lead pacing as LV tip to right ventricular (RV) lead coil to potentially improve pacing thresholds. However, anodal stimulation can result in unintentional RV pacing (anode) instead of LV pacing (cathode).

Methods

Patients enrolled in our center''s CRT registry had an echocardiogram, 6-minute walk (6MW), and Minnesota Living with HF Questionnaire (MLHFQ) pre-implant and 6 months after CRT. Electrocardiograms (12 lead) during RV, LV, and biventricular (BiV) pacing were obtained at the end of the implant in 102 patients. Anodal stimulation was defined as LV pacing QRS morphology on EKG being identical to RV pacing or consistent with fusion with RV and LV electrode capture. LV end systolic volume (LVESV) was measured by echo biplane Simpson''s method and CRT responder was defined as 15% or greater reduction in LVESV.

Results

Of the 102 patients, 46 (45.1%) had the final LV lead pacing configuration programmed LV (tip or ring) to RV (coil or ring). 3 of the 46 subjects (6.5%) had EKG findings consistent with anodal stimulation, not corrected intraoperatively. All anodal stimulation patients were nonresponders to CRT by echo criteria (reduction in LVESV 13.3 ± 0.6%, increase in EF 5.0 ± 1.4%) compared to 46% responders for those without anodal stimulation, (change in LVESV 18.7 ± 25.6%, EF 7.6 ±10.9%). None of the anodal stimulation patients were responders for the 6 minute walk, compared to 32 of 66 (48%) of those without anodal stimulation.

Conclusion

Anodal stimulation is a potential underrecognized and ameliorable cause of poor response to CRT.  相似文献   

2.
3.

Background

Cardiac Resynchronization Therapy (CRT) is indicated for the treatment of advanced heart failure with severe systolic dysfunction and intraventricular conduction delay. Patient selection for this technology is vital, though it remains unclear which patients benefit most from CRT. We tested the hypothesis that patients with non-ischemic cardiomyopathy have a superior mortality benefit from CRT than ischemic cardiomyopathy patients.

Methods

We evaluated 95 CRT patients to determine which factors predict mortality.

Results

Patients with non-ischemic cardiomyopathy had a significantly better prognosis than patients with ischemic cardiomyopathy.

Conclusion

Larger prospective studies can substantiate this finding and better delineate which patients benefit most from CRT.  相似文献   

4.

Background

Delayed left ventricular (LV) lateral wall activation is considered the electrical substrate that characterises patients suitable for cardiac resynchronisation therapy (CRT). Although typically associated with left bundle branch block, delayed LV lateral wall activation may also be present in patients with non-specific intraventricular conduction delay (IVCD). We assessed LV lateral wall activation in a cohort of CRT candidates with IVCD using coronary venous electroanatomical mapping, and investigated whether baseline QRS characteristics on the ECG can identify delayed LV lateral wall activation in this group of patients.

Methods

Twenty-three consecutive CRT candidates with IVCD underwent intra-procedural coronary venous electroanatomical mapping using EnSite NavX. Electrical activation time was measured in milliseconds from QRS onset and expressed as percentage of QRS duration. LV lateral wall activation was considered delayed if maximal activation time measured at the LV lateral wall (LVLW-AT) exceeded 75 % of the QRS duration. QRS morphology, duration, fragmentation, axis deviation, and left anterior/posterior fascicular block were assessed on baseline ECGs.

Results

Delayed LV lateral wall activation occurred in 12/23 patients (maximal LVLW-AT = 133 ± 20 ms [83 ± 5 % of QRS duration]). In these patients, the latest activated region was consistently located on the basal lateral wall. QRS duration, and prevalence of QRS fragmentation and left/right axis deviation, and left anterior/posterior fascicular block did not differ between patients with and without delayed LV lateral wall activation.

Conclusion

Coronary venous electroanatomical mapping can be used at the time of CRT implantation to determine the presence of delayed LV lateral wall activation in patients with IVCD. QRS characteristics on the ECG seem unable to identify delayed LV lateral wall activation in this subgroup of patients.  相似文献   

5.

Introduction

Non response to cardiac resynchronisation therapy (CRT) may be related to the position of the coronary sinus lead.

Methods

We studied the acute haemodynamic response (AHR) from alternative left ventricular (LV) endocardial pacing sites in clinical non-responders to CRT. AHR and the interval from QRS onset to LV sensing (Q-LV interval) from four different endocardial pacing sites were evaluated in 24 clinical non-responders. A rise in LVdP/dtmax ≥ 15 % from baseline was considered a positive AHR. We also compared the AHR from endocardial with the corresponding epicardial lead position.

Results

The implanted system showed an AHR ≥ 15 % in 5 patients. In 9 of the 19 remaining patients, AHR could be elevated to ≥ 15 % by endocardial LV pacing. The optimal endocardial pacing site was posterolateral. There was no significant difference in AHR between the epicardial and the corresponding endocardial position. The longest Q-LV interval corresponded with the best AHR in 12 out of the 14 patients with a positive AHR, with an average Q-LV/QRS width ratio of 90 %.

Conclusions

Acute haemodynamic testing may indicate an alternative endocardial pacing site with a positive AHR in clinical non-responders. The Q-LV interval is a strongly correlated with the optimal endocardial pacing site. Endocardial pacing opposite epicardial sites does not result in a better AHR.  相似文献   

6.

Objectives

The beneficial effects of cardiac resynchronization therapy (CRT) are thought to result from favorable left ventricular (LV) reverse remodeling, however CRT is only successful in about 70% of patients. Whether response to CRT is associated with a decrease in ventricular arrhythmias (VA) is still discussed controversially. Therefore, we investigated the incidence of VA in CRT responders in comparison with non-responders.

Methods

In this nonrandomized, two-center, observational study patients with moderate-to-severe heart failure, LV ejection fraction (LVEF) ≤35%, and QRS duration >120 ms undergoing CRT were included. After 6 months patients were classified as CRT responders or non-responders. Incidence of VA was compared between both groups by Kaplan-Meier analysis and Cox regression analysis. ROC analysis was performed to determine the aptitude of LVEF cut-off values to predict VA.

Results

In total 126 consecutive patients (64±11years; 67%male) were included, 74 were classified as responders and 52 as non-responders. While the mean LVEF at baseline was comparable in both groups (25±7% vs. 24±8%; P = 0.4583) only the responder group showed an improvement of LVEF (36±6% vs. 24±7; p<0.0001) under CRT. In total in 56 patients VA were observed during a mean follow-up of 28±14 months, with CRT responders experiencing fewer VA than non-responders (35% vs. 58%, p<0.0061). Secondary preventive CRT implantation was associated with a higher likelihood of VA. As determined by ROC analysis an increase of LVEF by >7% was found to be a predictor of a significantly lower incidence of VA (AUC = 0.606).

Conclusions

Improvement of left ventricular function under cardiac resynchronization therapy goes along with a reduced incidence of ventricular arrhythmia.  相似文献   

7.

Objective

To describe three DBS cases which presented with new side effects or loss of benefit from stimulation after long-term follow-up and to discuss the potential contributing factors.

Methods

A University of Florida (UF) database (INFORM) search was performed, identifying three patients, two Parkinson''s disease (PD) and one Essential Tremor (ET), with an unexpected change in long-term programming thresholds as compared to initial evaluation. Clinical follow-up, programming, imaging studies, and lead measurements were reviewed. The UF Institutional Review Board (IRB) approved this study.

Results

A substantial increase in the 3rd ventricular width (120%), Evans index (6%), ventricular index (5%), and cella media index (17%) was uncovered. A change in thresholds across lead contacts with a decrease in current densities as well as a relative lateral change of lead location was also observed. Hardware-related complications, lead migration, and impedance variability were not identified.

Conclusions

Potential factors contributing to long-term side effects should be examined during a DBS troubleshooting assessment. Clinicians should be aware that in DBS therapy there is delivery of electricity to a changing brain, and atrophy may possibly affect DBS programming settings as part of long-term follow-up.  相似文献   

8.
9.

Background

Predischarge defibrillation threshold testing is often performed a few days after ICD implantation in order to validate defibrillation thresholds obtained at the time of implant. Ventricular fibrillation is induced with such testing and causes an increase in serum Brain Natriuretic Peptide (BNP) levels. BNP is an indicator for cardiac stress. We wanted to examine the feasibility to alter the trend of BNP after predischarge testing in VVI, DDD and CRT ICD''s.

Methods

We measured BNP before predischarge testing and 5, 10, 20 and 40 minutes after predischarge testing in 13 groups with each 20 patients. We evaluated patients without post shock pacing and patients with a post shock pacing frequency of 60, 70, 80, 90 and 100 bpm and a duration of 30 and 60 sec as well as a post shock pacing frequency of 80 and 90 bpm and a duration of 120 sec post shock pacing.

Results

Patients without post shock pacing showed the highest BNP during the follow-up. The percentage values of BNP increased consistent significantly after 5 minutes compared with BNP before predischarge testing. The percentage values of BNP trend was significantly lower with a post shock pacing of 90 bpm and duration of 60 sec. In addition, we excluded a cardiac necrosis by predischarge testing because of similar values of myoglobin, cardiac troponin I and creatine kinase during the follow-up.

Conclusions

Our results suggested that post shock pacing with 90 bpm and duration of 60 sec as the best optimized post shock pacing frequency and duration for VVI, DDD and CRT ICD''s. A reduction of cardiac stress is going to be achieved with the optimization of the post shock pacing frequency and duration.  相似文献   

10.

Background

Super-responders to cardiac resynchronisation therapy (CRT) show an exceptional improvement in left ventricular ejection fraction (LVEF). Previous studies showed that apical rocking was independently associated with echocardiographic response to CRT. However, little is known about the association between apical rocking and super-response to CRT.

Objectives

To determine the independent association of LV apical rocking with super-response to CRT in a large cohort.

Methods

A cohort of 297 consecutive heart failure patients treated with primary indication for CRT-D were included in an observational registry. Apical rocking was defined as motion of the left ventricular (LV) apical myocardium perpendicular to the LV long axis. ‘Super-response’ was defined by the top quartile of LVEF response based on change from baseline to follow-up echocardiogram. Best-subset regression analysis identified predictors of LVEF super-response to CRT.

Results

Apical rocking was present in 45 % of patients. Super-responders had an absolute mean LVEF increase of 27 % (LVEF 22.0 % ± 5.7 at baseline and 49.0 % ± 7.5 at follow-up). Apical rocking was significantly more common in super-responders compared with non-super-responders (76 and 34 %, P < 0.001). In univariate analysis, female gender (OR 2.39, 95 % CI 1.38–4.11), lower LVEF at baseline (OR 0.91 95 % CI 0.87–0.95), non-ischaemic aetiology (OR 4.15, 95 % CI 2.33–7.39) and apical rocking (OR 6.19, 95 % CI 3.40–11.25) were associated with super-response. In multivariate analysis, apical rocking was still strongly associated with super-response (OR 5.82, 95 % CI 2.68–12.61). Super-responders showed an excellent clinical prognosis with a very low incidence of heart failure admission, cardiac mortality and appropriate ICD therapy.

Conclusion

Apical rocking is independently associated with super-response to CRT.

Electronic supplementary material

The online version of this article (doi:10.1007/s12471-015-0768-4) contains supplementary material, which is available to authorized users.  相似文献   

11.

Background

Cardiac resynchronization therapy (CRT) is indicated in symptomatic heart failure (HF) patients after achieving optimal medical therapy. However, there are still a large percentage of patients who do not respond to CRT. Malnutrition is a frequent comorbidity in patients with HF, and it is associated with a poorer prognosis. Here, we evaluate the nutritional status of patients assessed by Controlling Nutritional Status (CONUT) score and its association with structural remodeling and cardiovascular events.

Methods

We investigated the effect of CONUT on HF/death in 302 consecutive patients with a CRT device implanted between 2005 and 2015 in a single tertiary center. We categorized the patients into three groups: normal nutritional status (CONUT 0–1), mild malnutrition (CONUT 2–4) and moderate-severe malnutrition (CONUT?≥?5). Changes in nutritional status were assessed in patients with mild-to-severe malnutrition prior to CRT.

Results

One hundred and forty-eight patients exhibited normal nutritional status (49.0%), 99 patients exhibited mild malnutrition (32.8%) and 55 patients exhibited moderate-severe malnutrition (18.2%). CONUT scores of at least 2 were associated with higher risk of HF/death compared with CONUT 0–1. Significant left ventricular (LV) reverse remodeling was noted in patients with better nutritional status. In addition, those malnutrition patients at baseline that improved nutritional state exhibited fewer HF/death events at follow-up.

Conclusion

CONUT score prior to CRT was an independent risk factor of death/HF and was correlated with LV reverse remodeling. Improvements in CONUT score during long-term follow-up were associated with a reduction in the rate of HF/death.  相似文献   

12.

Background

The number of cardiac rhythm device implantations has been growing fast due to expanding indications and ageing of the population. Complications of implantation were rare in the trials. However, these involved small numbers and selected patients. Prospective real-life data are necessary to assess cardiac device implantation procedure-related risks.

Objective

To determine the incidence and predictors of lead-related re-intervention in a Dutch high-volume teaching hospital.

Methods

Data from all patients who underwent cardiac rhythm device implantation between January 2010 and December 2011 were collected in a prospective registry. At least 1 year of follow-up regarding re-intervention was available for all patients. Lead-related reasons for re-intervention were categorised into lead dislodgement, malfunctioning or perforation.

Results

One thousand nine hundred twenty-nine devices including 3909 leads were implanted. In 595 patients (30.8 %) a CRT-D/P was implanted. Lead-related re-intervention was necessary in 86 (4.4 %) patients; it was more common in younger and male patients, and due to either lead dislodgement (66 %), malfunctioning (20 %) or perforation (18 %). Coronary sinus lead dislodgement or malfunctioning was 1.4 %. Right atrial dislodgement (1.9 %, p < 0.001) or ICD lead dislodgement (1.8 %, p = 0.002) was more common than right ventricular dislodgement (0.3 %). The incidence of lead malfunctioning was higher (0.8 %) in ICD leads. An apical position of the right ventricular lead and lateral wall position of the right atrial lead were related to cardiac perforation.

Conclusions

The incidence of lead-related re-intervention was comparable with the literature. The majority of re-interventions were due to lead dislodgements, particularly with right atrial and ICD leads. Re-intervention due to coronary sinus lead dislodgement was rare.  相似文献   

13.

Background

The prognostic significance of paced QRS complex morphology on surface ECG remains unclear. This study aimed to assess long-term outcomes associated with variations in the paced QRS complex.

Methods

Adult patients who underwent dual-chamber pacemaker implantation with 20% or more ventricular pacing and a 12-lead ECG showing a paced complex were included. The paced QRS was analyzed in leads I and aVL. Long-term clinical and echocardiographic outcomes were compared at 5 years.

Results

The study included 844 patients (43.1% female; age 75.0?±?12.1). Patients with a longer paced QRS (pQRS) duration in lead I had a lower rate of atrial fibrillation (HR 0.80; p?=?0.03) and higher rate of systolic dysfunction (HR 1.17; p?<?0.001). Total pacing complex (TPC) duration was linked to higher rates of ICD implantation (HR 1.18; p?=?0.04) and systolic dysfunction (HR 1.22, p?<?0.001). Longer paced intrinsicoid deflection (pID) was associated with less atrial fibrillation (HR 0.75; p?=?0.01), more systolic dysfunction (HR 1.17; p?<?0.001), ICD implantation (HR 1.23; p?=?0.04), and CRT upgrade (HR 1.23; p?=?0.03). Exceeding thresholds for TPC, pQRS, and pID of 170, 146, and 112?ms in lead I, respectively, was associated with a substantial increase in systolic dysfunction over 5 years (p?<?0.001).

Conclusions

Longer durations of all tested parameters in lead I were associated with increased rates of left ventricular systolic dysfunction. ICD implantation and CRT upgrade were also linked to increased TPC and pID durations. Paradoxically, patients with longer pID and pQRS had less incident atrial fibrillation.  相似文献   

14.

Objectives

The aim of this study was to compare conventional versus steerable catheter guided coronary sinus (CS) cannulation in patients with advanced heart failure undergoing cardiac resynchronization therapy (CRT).

Background

Steerable catheter guided coronary sinus cannulation could reduce fluoroscopy time and contrast medium use during CRT implantation.

Methods

176 consecutive patients with ischemic and non-ischemic heart failure undergoing CRT implantation from January 2008 to December 2012 at the University Hospital of Cologne were identified. During the study period two concurrent CS cannulation techniques were used: standard CS cannulation technique (standard-group, n = 113) and CS cannulation using a steerable electrophysiology (EP) catheter (EPCath-group, n = 63). Propensity-score matched pairs of conventional and EP-catheter guided CS cannulation made up the study population (n = 59 pairs). Primary endpoints were total fluoroscopy time and contrast medium amount used during procedure.

Results

The total fluoroscopy time was 30.9 min (interquartile range (IQR), 19.9–44.0 min) in the standard-group and 23.4 min (IQR, 14.2-34-2 min) in the EPCath-group (p = 0.011). More contrast medium was used in the standard-group (60.0 ml, IQR, 30.0–100 ml) compared to 25.0 ml (IQR, 20.0–50.0 ml) in the EPCath-group (P<0.001).

Conclusions

Use of steerable EP catheter was associated with significant reduction of fluoroscopy time and contrast medium use in patients undergoing CRT implantation.  相似文献   

15.

Background

Right ventricular (RV) dysfunction has been associated with poor prognosis in chronic heart failure (HF). However, less data is available about the role of RV dysfunction in patients with cardiac resynchronization therapy (CRT). We aimed to investigate if RV dysfunction would predict outcome in CRT.

Design

We enrolled prospectively ninety-three consecutive HF patients in this single center observational study. All patients underwent clinical evaluation and echocardiography before CRT and 6 months after implantation. We assessed RV geometry and function by using speckle tracking imaging and calculated strain parameters. We performed multivariable Cox regression models to test mortality at 6 months and at 24 months.

Results

RV dysfunction, characterized by decreased RVGLS (RV global longitudinal strain) [10.2 (7.0–12.8) vs. 19.5 (15.0–23.9) %, p<0.0001] and RVFWS (RV free wall strain) [15.6 (10.0–19.3) vs. 17.4 (10.5–22.2) %, p = 0.04], improved 6 months after CRT implantation. Increasing baseline RVGLS and RVFWS predicted survival independent of other parameters at 6 months [hazard ratio (HR) = 0.37 (0.15–0.90), p = 0.02 and HR = 0.42 (0.19–0.89), p = 0.02; per 1 standard deviation increase, respectively]. RVGLS proved to be a significant independent predictor of mortality at 24 months [HR = 0.53 (0.32–0.86), p = 0.01], and RVFWS showed a strong tendency [HR = 0.64 (0.40–1.00), p = 0.05]. The 24-month survival was significantly impaired in patients with RVGLS below 10.04% before CRT implantation [area under the curve = 0.72 (0.60–0.84), p = 0.002, log-rank p = 0.0008; HR = 5.23 (1.76–15.48), p = 0.003].

Conclusions

Our findings indicate that baseline RV dysfunction is associated with poor short-term and long-term prognosis after CRT implantation.  相似文献   

16.

Introduction

Phrenic nerve stimulation (PNS) could be prevented by a silastic patch over the epicardial lead. We studied the effects in preventing PNS by placing a silastic patch directly over an epicardial lead or placing a graft around the phrenic nerve (PN).

Methods and Results

Fourteen Lanyu swine were enrolled. A bipolar lead was placed epicardially on the left ventricle (LV) inferior to the PN. An implantable cardioverter-defibrillator (ICD) lead was placed into the right ventricle (RV). The maximal influential distance (MID) was measured under 3 pacing configurations to express the influential electrical field on the PN. The threshold of the LV and PN were evaluated epicardially. Then, PTFE patches of different sizes (10×10 mm, 20×20 mm and 30×30 mm) were placed between the LV lead and PN to study the rise in PN threshold in 7 swine. On the other hand, the PN were surrounded by a PTFE graft of different lengths (10 mm, 20 mm, and 30 mm) in the remaining 7 swine. LV-bipolar pacing showed the shortest MID when compared to the other 2 unipolar pacing configurations at pacing voltage of 10 V. The patch was most effective in preventing PNS during LV-bipolar pacing. PNS was prevented under all circumstances with a larger PTFE patch (30×30 mm) or long graft (30 mm).

Conclusions

PNS was avoided by placing a PTFE patch over the LV lead or a graft around the PN despite pacing configurations. Hence if PNS persisted during CRT implantation, a PTFE patch on the LV lead or a graft around the PN could be considered.  相似文献   

17.

Background

Patient-reported factors have largely been neglected in search of predictors of response to cardiac resynchronisation therapy (CRT). The current study aimed to examine the independent value of pre-implantation patient-reported health status in predicting four-year survival and cardiac-related hospitalisation of CRT patients.

Methods

Consecutive patients (N = 139) indicated to receive a first-time CRT-defibrillator at the University Medical Center Utrecht were asked to complete a set of questionnaires prior to implantation. The Kansas City Cardiomyopathy Questionnaire (KCCQ) was used to assess heart failure-specific health status. Data on patients’ demographic, clinical and psychological characteristics at baseline, and on cardiac-related hospitalisations and all-cause deaths during a median follow-up of 3.9 years were obtained from purpose-designed questionnaires and patients’ medical records.

Results

Results of multivariable Cox regression analyses showed that poor patient-reported health status (KCCQ score < 50) prior to implantation was associated with a 2.5-fold increased risk of cardiac hospitalisation or all-cause death, independent of sociodemographic, clinical and psychological risk factors (adjusted hazard ratio 2.46, 95 % confidence interval (CI) 1.30–4.65). Poor health status was not significantly associated with the absolute number of cardiac-related hospital admissions, but with the total number of days spent in hospital during follow-up (adjusted incidence rate ratio 3.20, 95 % CI 1.88–5.44).

Conclusions

Patient-reported health status assessed prior to CRT identifies patients at risk for poor survival and prolonged hospital stays, independent of traditional risk factors. These results emphasise the importance of incorporating health status measures in cardiovascular research and patient management. Heart failure patients reporting poor health status should be identified and offered appropriate additional treatment programs.  相似文献   

18.

Background

In patients with left ventricular (LV) dysssynchrony, contraction that doesn’t fall into ejection period (LVEj) results in a waste of energy due to inappropriate contraction timing, which was now widely treated by cardiac resynchronization therapy(CRT). Myocardial Contraction Efficiency was defined as the ratio of Efficient Contraction Time (ECTR) and amplitude of efficient contraction (ECR) during LVEj against that in the entire cardiac cycle. This study prospectively investigated whether efficiency indexes could predict CRT outcome.

Methods

Our prospective pilot study including 70 CRT candidates, parameters of myocardial contraction timing and contractility were measured by speckle tracking echocardiography (STE) and efficiency indexes were calculated accordingly at baseline and at 6-month follow-up. Primary outcome events were predefined as death or HF hospitalization, and secondary outcome events were defined as all-cause death during the follow-up. 16-segement Standard deviation of time to onset strain (TTO-16SD) and time to peak strain (TTP-16SD) were included as the dyssynchrony indexes.

Results

According to LV end systolic volume (LVESV) and LV eject fraction(LVEF) values at 6-month follow-up, subjects were classified into responder and non-responder groups, ECR (OR 0.87, 95%CI 0.78–0.97, P?<?0.05) and maximum longitudinal strain (MLS) (OR 2.22, 95%CI 1.36–3.61, P?<?0.01) were the two independent predictors for CRT response, Both TTO-16SD and TTP-16SD failed to predict outcome. Patients with poorer myocardial contraction efficiency and better contractility are more likely to benefit from CRT.

Conclusions

STE can evaluate left ventricular contraction efficiency and contractility to predict CRT response. When analyzing myocardial strain by STE, contraction during LVEj should be highlighted.
  相似文献   

19.

Background

Good and reliable atrial sensing is a fundamental part of atrioventricular (AV) synchrony in dual chamber pacemakers. Due to the floating nature of atrial sensing electrode in single pass dual chamber pacemakers (VDD) compared with two-lead dual chamber pacemakers (DDD), they are more prone to atrial under-sensing and the resulting loss of AV synchrony. We hypothesized that there is a relation between the chronicity of AV block and the amplitude of intracardiac atrial signal amplitudes (IASA).

Methods

Detailed demographic, electrocardiographic and echocardiographic data were recorded in 34 consecutive patients with congenital and acquired complete heart block (CHB). The intracardiac atrial signal amplitudes (IASA) were recorded at implantation time, 48 hours and 2 months post-implantation and compared between the two groups of patients.

Results

The mean age of the study group was 38.73±12.53 years (congenital: 30.08±11.07, acquired: 47.38±6.5). There were no important differences in left atrial or ventricular sizes and in P-wave amplitude in lead II, but the IASA was significantly higher in the congenital group at implantation time (5.21±1.86 vs. 3.38±0.84 mV, P<0.001) and during the follow-up.

Conclusion

The intracardiac atrial signal amplitudes were higher in congenital CHB compared with the acquired CHB. Chronicity (and may be the congenital type) of CHB may be an affecting factor in case selection for VDD pacemaker implantation.  相似文献   

20.

Background

Change in left ventricular end-systolic volume (∆LVESV) is the most frequently used surrogate marker in measuring response to cardiac resynchronisation therapy (CRT). We investigated whether ∆LVESV is the best measure to discriminate between a favourable and unfavourable outcome and whether this is equally applicable to non-ischaemic and ischaemic cardiomyopathy.

Methods

205 CRT patients (age 65 ± 12 years, 69 % men) were included. At baseline and 6 months echocardiographic studies, exercise testing and laboratory measurements were performed. CRT response was assessed by: ∆LVESV, ∆LV ejection fraction (LVEF), ∆ interventricular mechanical delay, ∆VO2 peak, ∆VE/VCO2, ∆BNP, ∆creatinine, ∆NYHA, and ∆QRS. These were correlated to the occurrence of major adverse cardiac events (MACE) between 6 and 24 months.

Results

MACE occurred in 19 % of the patients (non-ischaemic: 13 %, ischaemic: 24 %). ∆LVESV remained the only surrogate marker for CRT response for the total population and patients with non-ischaemic cardiomyopathy, showing areas under the curve (AUC) of 0.69 and 0.850, respectively. For ischaemic cardiomyopathy, ∆BNP was the best surrogate marker showing an AUC of 0.66.

Conclusion

∆LVESV is an excellent surrogate marker measuring CRT response concerning long-term outcome for non-ischaemic cardiomyopathy. ∆LVESV is not suitable for ischaemic cardiomyopathy in which measuring CRT response remains difficult.  相似文献   