首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到20条相似文献,搜索用时 15 毫秒
1.
PurposeImage-guided radiotherapy (IGRT) based on bone matching can produce large target-positioning errors because of expiration breath-hold reproducibility during stereotactic body radiation therapy (SBRT) for liver tumors. Therefore, the feasibility of diaphragm-based 3D image matching between planning computed tomography (CT) and pretreatment cone-beam CT was investigated.MethodsIn 59 liver SBRT cases, Lipiodol uptake after transarterial chemoembolization was defined as a tumor marker. Further, the relative isocenter coordinate that was obtained by Lipiodol matching was defined as the reference coordinate. The distance between the relative isocenter coordinate and reference coordinate, which was obtained from diaphragm matching and bone matching techniques, was defined as the target positioning error. Furthermore, the target positioning error between liver matching and Lipiodol matching was evaluated.ResultsThe positioning errors in all directions by the diaphragm matching were significantly smaller than those obtained by using by the bone matching technique (p < 0.05). Further, the positioning errors in the A-P and C-C directions that were obtained by using liver matching were significantly smaller than those obtained by using bone matching (p < 0.05). The estimated PTV margins calculated by the formula proposed by van Herk for diaphragm matching, liver matching, and bone matching were 5.0 mm, 5.0 mm, and 11.6 mm in the C-C direction; 3.6 mm, 2.4 mm, and 6.9 mm in the A-P direction; and 2.6 mm, 4.1 mm, and 4.6 mm in the L-R direction, respectively.ConclusionsDiaphragm matching-based IGRT may be an alternative image matching technique for determining liver tumor positions in patients.  相似文献   

2.
3.
Purpose/objectiveStereotactic ablative body radiotherapy (SABR) in multi-centre trials requires rigorous quality assurance to ensure safe and consistent treatment for all trial participants. We report results of vertebral SABR dosimetry credentialing for the ALTG/TROG NIVORAD trial.Material/methodsCentres with a previous SABR site visit performed axial film measurement of the benchmarking vertebral plan in a local phantom and submitted radiochromic film images for analysis. Remaining centres had on-site review of SABR processes and axial film measurement of the vertebral benchmarking plan. Films were analysed for dosimetric and positional accuracy: gamma analysis (>90% passing 2%/2mm/10% threshold) and ≤ 1 mm positional accuracy at target-cord interface was required.Results19 centres were credentialed; 11 had on-site measurement. Delivery devices included linear accelerator, TomoTherapy and CyberKnife systems. Five centres did not achieve 90% gamma passing rate. Of these, three were out of tolerance (OOT) in low (<5Gy) dose regions and > 80% passing rate and deemed acceptable. Two were OOT over the full dose range: one elected not to remeasure; the other also had positional discrepancy greater than 1 mm and repeat measurement with a new plan was in tolerance. The original OOT was attributed to inappropriate MLC constraints. All centres delivered planned target-cord dose gradient within 1 mm.ConclusionCredentialing measurements for vertebral SABR in a multi-centre trial showed although the majority of centres delivered accurate vertebral SABR, there is high value in independent audit measurements. One centre with inappropriate MLC settings was detected, which may have resulted in delivery of clinically unacceptable vertebral SABR plans.  相似文献   

4.
Background and PurposeWith the increasingly prominent role of stereotactic radiosurgery in radiation therapy, there is a clinical need for robust, efficient, and accurate solutions for targeting multiple sites with one patient setup. The end-to-end accuracy of high definition dynamic radiosurgery with Elekta treatment planning and delivery systems was investigated in this study.Materials and MethodsA patient-derived CT scan was used to create a radiosurgery plan to seven targets in the brain. Monaco was used for treatment planning using 5 VMAT non-coplanar arcs. Prior to delivery, 3D-printed phantoms from RTsafe were ordered including a gel phantom for 3D dosimetry, phantom with 2D film insert, and an ion chamber phantom for point dose measurement. Delivery was performed using the Elekta VersaHD, XVI cone-beam CT, and HexaPOD six degree of freedom tabletop.ResultsAbsolute dose accuracy was verified within 2%. 3D global gamma analysis in the film measurement revealed 3%/2 mm passing rates >95%. Gel dosimetry 3D global gamma analysis (3%/2 mm) were above 90% for all targets with the exception of one. Results were indicative of typical end-to-end accuracies (<1 mm spatial uncertainty, 2% dose accuracy) within 4 cm of isocenter. Beyond 4 cm, 2 mm accuracy was found.ConclusionsHigh definition dynamic radiosurgery expands clinically acceptable stereotactic accuracy to a sphere around isocenter allowing for radiosurgery of several targets with one setup with a high degree of dosimetric precision. Gel dosimetry proved to be an essential tool for the validation of the 3D dose distributions in this technique.  相似文献   

5.
PurposeTo investigate the use of dual isocenters for VMAT planning in patients with lymph node positive synchronous bilateral breast cancer (BBC) compared to a single isocenter option.MethodsTreatment plans of 11 patients with lymph node positive BBC were retrospectively analyzed using two different VMAT planning techniques: dual-isocenter split-arc VMAT plans (Iso2) were compared with mono-isocenter VMAT plans (Iso1). For Iso2 plans, PTV dose was investigated after introducing ±2 and ±5 mm couch shift errors between the two isocenters in the lateral, longitudinal and vertical direction.ResultsFor both techniques the planning aims for PTV coverage and OARs were met. The mean dose for the bilateral lungs and heart was reduced from 11.3 Gy and 3.8 Gy to 10.9 Gy (p < .05) and 3.6 Gy (p < .05), respectively, for Iso2 plans when compared to Iso1 plans. Positive statistically significant correlation (rho = 0.76, p = .006) was found between PTV volume and D2ccPTV for Iso1 plans. No clinically significant change was seen in the D98CTV or D2ccPTV after the 2 and 5 mm errors were introduced between isocenters for Iso2 plans.ConclusionsThe split arc method was shown to be a feasible treatment technique in the case of synchronous BBC for both mono and dual isocenter techniques. The dose parameters were slightly favoring dual-isocenter option instead of mono-isocenter. The dual-isocenter method was shown to be a robust treatment option in the presence of ≤5 mm errors in the shifts between the two isocenters.  相似文献   

6.
We developed an efficient postal audit system to independently assess the delivered dose using radiophotoluminescent glass dosimeters (RPLDs) and the positional differences of fields using EBT3 film at the axial plane for intensity-modulated radiotherapy (IMRT). The audit phantom had a C-shaped target structure as a planning target volume (PTV) with four measurement points for the RPLDs and a cylindrical structure as the organ at risk (OAR) for one measurement point. The phantoms were sent to 24 institutions. Point dose measurements with a 0.6 cm3 PTW farmer chamber were also performed to justify glass dosimetry in IMRT. The measured dose with the RPLDs was compared to the calculated dose in the institution’s treatment planning system (TPS). The mean ± 1.96σ of the ratio of the measured dose with the RPLDs to the farmer chamber was 0.997 ± 0.024 with no significant difference (p = .175). The investigations demonstrated that glass dosimetry was reliable with a high measurement accuracy comparable to the chamber. The mean ± 1.96σ for the dose differences with a reference of the TPS dose for the PTV and the OAR was 0.1 ± 2.5% and −2.1 ± 17.8%, respectively. The mean ± 1.96σ for the right-left and the anterior–posterior direction was −0.9 ± 2.8 and 0.5 ± 1.4 mm, respectively. This study is the first report to justify glass dosimetry for implementation in IMRT audit in Japan. We demonstrate that our postal audit system has high accuracy with a high-level criterion of 3%/3 mm.  相似文献   

7.
PurposeA dosimetric audit of Ir-192 high dose rate (HDR) brachytherapy remote after-loading units was carried out in 2019. All six brachytherapy departments on the island of Ireland participated in an end-to-end test and in a review of local HDR dosimetry procedures.Materials and methodsA 3D-printed customised phantom was created to position the following detectors at known distances from the HDR source: a Farmer ionization chamber, GafChromic film and thermoluminescent dosimeters (TLDs). Dedicated HDR applicator needles were used to position an Ir-192 source at 2 cm distance from these detectors. The end-to-end dosimetry audit pathway was performed at each host site and included the stages of imaging, applicator reconstruction, treatment planning and delivery. Deviations between planned and measured dose distributions were quantified using gamma analysis methods. Local procedures were also discussed between auditors and hosts.ResultsThe mean difference between Reference Air Kerma Rate (RAKR) measured during the audit and RAKR specified by the vendor source certificate was 1.3%. The results of end-to-end tests showed a mean difference between calculated and measured dose of 2.5% with TLDs and less than 0.5% with Farmer chamber measurements. GafChromic films showed a mean gamma passing rates of >95% for plastic and metal applicators with 2%/1 mm global tolerance criteria.ConclusionsThe results of this audit indicate dosimetric consistency between centres. The ‘end to end’ dosimetry audit methodology for HDR brachytherapy has been successfully implemented in a multicentre environment, which included different models of Ir-192 sources and different treatment planning systems.The ability to create a 3D-printed water-equivalent phantom customised to accurately position all three detector types simultaneously at controlled distances from the Ir-192 source under evaluation gives good reproducibility for end-to-end methodology.  相似文献   

8.
Background and purposeThe Medical Physics Division of the Portuguese Physics Society (DFM_SPF) in collaboration with the IAEA, carried out a national auditing project in radiotherapy, between September 2011 and April 2012. The objective of this audit was to ensure the optimal usage of treatment planning systems. The national results are presented in this paper.Material and methodsThe audit methodology simulated all steps of external beam radiotherapy workflow, from image acquisition to treatment planning and dose delivery. A thorax CIRS phantom lend by IAEA was used in 8 planning test-cases for photon beams corresponding to 15 measuring points (33 point dose results, including individual fields in multi-field test cases and 5 sum results) in different phantom materials covering a set of typical clinical delivery techniques in 3D Conformal Radiotherapy.ResultsAll 24 radiotherapy centers in Portugal have participated. 50 photon beams with energies 4–18 MV have been audited using 25 linear accelerators and 32 calculation algorithms.In general a very good consistency was observed for the same type of algorithm in all centres and for each beam quality.ConclusionsThe overall results confirmed that the national status of TPS calculations and dose delivery for 3D conformal radiotherapy is generally acceptable with no major causes for concern. This project contributed to the strengthening of the cooperation between the centres and professionals, paving the way to further national collaborations.  相似文献   

9.
PurposeCurrent quality assurance of radiotherapy involving bony regions generally utilises homogeneous phantoms and dose calculations, ignoring the challenges of heterogeneities with dosimetry problems likely occurring around bone. Anthropomorphic phantoms with synthetic bony materials enable realistic end-to-end testing in clinical scenarios. This work reports on measurements and calculated corrections required to directly report dose in bony materials in the context of comprehensive end-to-end dosimetry audit measurements (63 plans, 6 planning systems).Materials and methodsRadiochromic film and microDiamond measurements were performed in an anthropomorphic spine phantom containing bone equivalent materials. Medium dependent correction factors, kmed, were established using 6 MV and 10 MV Linear Accelerator Monte Carlo simulations to account for the detectors being calibrated in water, but measuring in regions of bony material. Both cortical and trabecular bony material were investigated for verification of dose calculations in dose-to-medium (Dm,m) and dose-to-water (Dw,w) scenarios.ResultsFor Dm,m calculations, modelled correction factors for cortical and trabecular bone in film measurements, and for trabecular bone in microDiamond measurements were 0.875(±0.1%), 0.953(±0.3%) and 0.962(±0.4%), respectively. For Dw,w calculations, the corrections were 0.920(±0.1%), 0.982(±0.3%) and 0.993(±0.4%), respectively. In the audit, application of the correction factors improves the mean agreement between treatment plans and measured microDiamond dose from −2.4%(±3.9%) to 0.4%(±3.7%).ConclusionMonte Carlo simulations provide a method for correcting the dose measured in bony materials allowing more accurate comparison with treatment planning system doses. In verification measurements, algorithm specific correction factors should be applied to account for variations in bony material for calculations based on Dm,m and Dw,w.  相似文献   

10.
摘要 目的:研究表皮生长因子受体(EGFR)突变对替莫唑胺联合图像引导大分割放射(IGRT)治疗非小细胞肺癌脑转移瘤临床疗效的影响。方法:选择2015年1月到2018年12月在我院接受治疗的非小细胞肺癌脑转移患者86例,根据是否出现EGFR突变分为对照组(EGFR未突变组)和研究组(EGFR突变组),每组43人,两组患者均接受替莫唑胺联合IGRT大分割放射治疗。比较两组患者临床治疗疗效、不良反应发生情况、复发时间、生存时间和生活质量。结果:研究组患者临床治疗总有效率较对照组患者高(P<0.05)。研究组患者治疗后复发时间和生存时间均显著高于对照组患者(P<0.05)。两组患者治疗期间头痛、恶心、疲乏以及神经毒性等不良反应的发生情况比较无显著差异(P>0.05)。两组患者治疗前生活质量KarnofSky活动状态评分(KPS)和肺癌相关症状量表(LCSS评分)无显著差异(P>0.05);治疗后,研究组患者KPS评分显著高于对照组(P<0.05),而LCSS评分显著低于对照组患者(P<0.05)。结论:替莫唑胺联合IGRT大分割放射治疗EGFR突变的非小细胞肺癌脑转移瘤临床疗效更好,并且治疗后患者生活质量更优。  相似文献   

11.
PurposeThis work aims to validate new 6D couch features and their implementation for seated radiotherapy in RayStation (RS) treatment planning system (TPS).Materials and methodsIn RS TPS, new 6D couch features are (i) chair support device, (ii) patient treatment option of “Sitting: face towards the front of the chair”, and (iii) patient support pitch and roll capabilities. The validation of pitch and roll was performed by comparing TPS generated DRRs with planar x-rays. Dosimetric tests through measurement by 2D ion chamber array were performed for beams created with varied scanning and treatment orientation and 6D couch rotations. For the implementation of 6D couch features for treatments in a seated position, the TPS and oncology information system (Mosaiq) settings are described for a commercial chair. An end-to-end test using an anthropomorphic phantom was performed to test the complete workflow from simulation to treatment delivery.ResultsThe 6D couch features were found to have a consistent implementation that met IEC 61712 standard. The DRRs were found to have an acceptable agreement with planar x-rays based on visual inspection. For dose map comparison between measured and calculated, the gamma index analysis for all the beams was >95% at a 3% dose-difference and 3 mm distance-to-agreement tolerances. For an end-to end-testing, the phantom was successfully set up at isocenter in the seated position and treatment was delivered.ConclusionsChair-based treatments in a seated position can be implemented in RayStation through the use of newly released 6D couch features.  相似文献   

12.
PurposeTo investigate and improve the domestic standard of radiation therapy in the Republic of Korea.MethodsOn-site audits were performed for 13 institutions in the Republic of Korea. Six items were investigated by on-site visits of each radiation therapy institution, including collimator, gantry, and couch rotation isocenter check; coincidence between light and radiation fields; photon beam flatness and symmetry; electron beam flatness and symmetry; physical wedge transmission factors; and photon beam and electron beam outputs.ResultsThe average deviations of mechanical collimator, gantry, and couch rotation isocenter were less than 1 mm. Those of radiation isocenter were also less than 1 mm. The average difference between light and radiation fields was 0.9 ± 0.6 mm for the field size of 20 cm × 20 cm. The average values of flatness and symmetry of the photon beams were 2.9% ± 0.6% and 1.1% ± 0.7%, respectively. Those of electron beams were 2.5% ± 0.7% and 0.6% ± 1.0%, respectively. Every institutions showed wedge transmission factor deviations less than 2% except one institution. The output deviations of both photon and electron beams were less than ±3% for every institution.ConclusionsThrough the on-site audit program, we could effectively detect an inappropriately operating linacs and provide some recommendations. The standard of radiation therapy in Korea is expected to improve through such on-site audits.  相似文献   

13.
PurposeIn radiotherapy, accurate calculation of patient radiation dose is very important for good clinical outcome. In the presence of metallic implants, the dose calculation accuracy could be compromised by metal artefacts generated in computed tomography (CT) images of patients. This study investigates the influence of metal-induced CT artefacts on MC dose calculations in a pelvic prosthesis phantom.MethodsA pelvic phantom containing unilateral Ti prosthesis was CT-scanned and accurate Hounsfield unit (HU) values were assigned to known materials of the phantom as opposed to HU values produced through the artefact CT images of the phantom. Using the DOSXYZnrc MC code, dose calculations were computed in the phantom model constructed from the original CT images containing the artefacts and artefact-free images made from the exact geometry of the phantom with known materials. The dose calculations were benchmarked against Gafchromic EBT3 film measurements using 15 MeV electron and 10 MV photon beams.ResultsThe average deviations between film and MC dose data decreased from 3 ± 2% to 1 ± 1% and from about 6 ± 2% to 3 ± 1% for the artefact and artefact-free phantom models against film data for the electron and photon fields, respectively.ConclusionsFor the Ti prosthesis phantom, the presence of metal-induced CT artefacts could cause dose inaccuracies of about 3%. Construction of an artefact-free phantom model made from the exact geometry of the phantom with known materials to overcome the effect of artefacts is advantageous compared to using CT data directly of which the exact tissue composition is not well-known.  相似文献   

14.
AimWe aim to evaluate the variables affecting the frequency of adaptive radiotherapy (ART) in vulvar cancer.BackgroundART may be needed throughout a definitive RT course for vulvar carcinoma due to changes in patient’s anatomy and tumor response.Materials and methodsCharts of patients charts who had been treated with definitive concurrent chemo-radiotherapy for vulvar carcinoma, between January 2015 and December 2019 were inquired. Radiation therapy was delivered using intensity modulated radiotherapy (IMRT) with daily image-guided radiotherapy (IGRT). ART was defined as re-simulation and re-planning based on deformation in the irradiated volume by more than 1 cm. Univariate analysis was conducted to study the impact of patient’s demographics as well as tumor characteristics on the frequency of ART.Results22 patients were eligible for analysis. Median age at diagnosis was 55 years (range 43–82). Radiotherapy dose was 60−66 Gy over 30–35 fractions (fx). Median primary tumor volume was 30cc (9–140). Median Body Mass Index (BMI) was 32 (range 21–40). Thirteen out of 22 patients (59%) required ART, with median timing at 25 fx (19–31). On univariate analysis, larger primary tumor volume (> = 30cc) was associated significantly with increased frequency of ART (p value = 0.0005). There was no significant impact of ART on the frequency with respect to patient’s age, BMI, tumor stage, grade and location.ConclusionChanges in radiation target volume are common among vulvar carcinoma patients who are treated with definitive radiotherapy, especially large primary tumors. This review highlights the importance of ART for patients with vulvar carcinoma treated with definitive radiotherapy.  相似文献   

15.
AimIn measuring exit fluences, there are several sources of deviations which include the changes in the entrance fluence, changes in the detector response and patient orientation or geometry. The purpose of this work is to quantify these sources of errors.BackgroundThe use of the volumetric modulated arc therapy treatment with the help of image guidance in radiotherapy results in high accuracy of delivering complex dose distributions while sparing critical organs. The transit dosimetry has the potential of Verifying dose delivery by the linac, Multileaf collimator positional accuracy and the calculation of dose to a patient or phantom.Materials and methodsThe quantification of errors caused by a machine delivery is done by comparing static and arc picket fence test for 30 days. A RapidArc plan, created for the pelvis site was delivered without and with Rando phantom and exit portal images were acquired. The day to day dose variation were analysed by comparing the daily exit dose images during the course of treatment. The gamma criterion used for analysis is 3% dose difference and 3 mm distance to agreement with a threshold of 10% of maximum dose.ResultsThe maximum standard deviation for the static and arc picket fence test fields were 0.19 CU and 1.3 CU, respectively. The delivery of the RapidArc plans without a phantom shows the maximum standard deviation of 1.85 CU and the maximum gamma value of 0.59. The maximum gamma value for the RapidArc plan delivered with the phantom was found to be 1.2. The largest observed fluence deviation during the delivery to patient was 5.7% and the maximum standard deviation was 4.1 CU.ConclusionIt is found from this study that the variation due to patient anatomy and interfraction organ motion is significant.  相似文献   

16.
PurposeTo develop and test the suitability and performance of a comprehensive quality assurance (QA) phantom for the Small Animal Radiation Research Platform (SARRP).Methods and materialsA QA phantom was developed for carrying out daily, monthly and annual QA tasks including: imaging, dosimetry and treatment planning system (TPS) performance evaluation of the SARRP. The QA phantom consists of 15 (60 × 60 × 5 mm3) kV-energy tissue equivalent solid water slabs. The phantom can incorporate optically stimulated luminescence dosimeters (OSLD), Mosfet or film. One slab, with inserts and another slab with hole patterns are particularly designed for image QA.ResultsOutput constancy measurement results showed daily variations within 3%. Using the Mosfet in phantom as target, results showed that the difference between TPS calculations and measurements was within 5%. Annual QA results for the Percentage depth dose (PDD) curves, lateral beam profiles, beam flatness and beam profile symmetry were found consistent with results obtained at commissioning. PDD curves obtained using film and OSLDs showed good agreement. Image QA was performed monthly, with image-quality parameters assessed in terms of CBCT image geometric accuracy, CT number accuracy, image spatial resolution, noise and image uniformity.ConclusionsThe results show that the developed QA phantom can be employed as a tool for comprehensive performance evaluation of the SARRP. The study provides a useful reference for development of a comprehensive quality assurance program for the SARRP and other similar small animal irradiators, with proposed tolerances and frequency of required tests.  相似文献   

17.
AimIn this study, an accuracy survey of intensity-modulated radiation therapy (IMRT) and volumetric arc radiation therapy (VMAT) implementation in radiotherapy centers in Thailand was conducted.BackgroundIt is well recognized that there is a need for radiotherapy centers to evaluate the accuracy levels of their current practices, and use the related information to identify opportunities for future development.Materials and methodsAn end-to-end test using a CIRS thorax phantom was carried out at 8 participating centers. Based on each center's protocol for simulation and planning, linac-based IMRT or VMAT plans were generated following the IAEA (CRP E24017) guidelines. Point doses in the region of PTVs and OARs were obtained from 5 ionization chamber readings and the dose distribution from the radiochromic films. The global gamma indices of the measurement doses and the treatment planning system calculation doses were compared.ResultsThe large majority of the RT centers (6/8) fulfilled the dosimetric goals, with the measured and calculated doses at the specification points agreeing within ±3% for PTV and ±5% for OARS. At 2 centers, TPS underestimated the lung doses by about 6% and spinal cord doses by 8%. The mean percentage gamma pass rates for the 8 centers were 98.29 ± 0.67% (for the 3%/3 mm criterion) and 96.72 ± 0.84% (for the 2%/2 mm criterion).ConclusionsThe 8 participating RT centers achieved a satisfactory quality level of IMRT/VMAT clinical implementation.  相似文献   

18.
PurposeTo investigate the effectiveness of an EPID-based 3D transit dosimetry system in detecting deliberately introduced errors during VMAT delivery.MethodsAn Alderson phantom was irradiated using four VMAT treatment plans (one prostate, two head-and-neck and one lung case) in which delivery, thickness and setup errors were introduced. EPID measurements were performed to reconstruct 3D dose distributions of “error” plans, which were compared with “no-error” plans using the mean gamma (γmean), near-maximum gamma (γ1%) and the difference in isocenter dose (ΔDisoc) as metrics.ResultsOut of a total of 42 serious errors, the number of errors detected was 33 (79%), and 27 out of 30 (90%) if setup errors are not included. The system was able to pick up errors of 5 mm movement of a leaf bank, a wrong collimator rotation angle and a wrong photon beam energy. A change in phantom thickness of 1 cm was detected for all cases, while only for the head-and-neck plans a 2 cm horizontal and vertical shift of the phantom were alerted. A single leaf error of 5 mm could be detected for the lung plan only.ConclusionAlthough performed for a limited number of cases and error types, this study shows that EPID-based 3D transit dosimetry is able to detect a number of serious errors in dose delivery, leaf bank position and patient thickness during VMAT delivery. Errors in patient setup and single leaf position can only be detected in specific cases.  相似文献   

19.
BackgroundCervical cancer is a public health problem in Latin America. Radiotherapy plays a fundamental role both as definitive or adjuvant treatment. There are important intra and inter-country differences regarding access and availability of radiotherapy facilities in this region. The aim of a study was to standardize the basic clinical and technical criteria for the radiation treatment of patients with CC in Chile and provide a guide for Latin American Radiation Oncologists.Materials and methodsForty-one expert radiation oncologists from the Chilean Radiation Oncology Society made a consensus using the Delphi methodology.ResultsThere was a high degree of agreement for each of the recommendations. Those with the lowest percentage were related to the definition of the conformal 3D technique as the standard for definitive external radiotherapy (81%) and the criteria for extended nodal irradiation (85%).ConclusionsThese recommendations present an updated guide for radiotherapy treatment of patients with cervical cancer for Latin America. Those should be implemented according to local resources of each institution.  相似文献   

20.
The demand for greater accuracy of intensity-modulated radiotherapy (IMRT) has driven the development of more advanced verification systems for image-guided radiotherapy (IGRT). The purpose of this study is to investigate setup discrepancies measured between an orthogonal X-ray guidance system (XGS-10) and cone-beam computed tomography (CBCT) of Varian in the IMRT of patients with nasopharyngeal cancer (NPC). The setup errors measured by XGS-10 and CBCT at the treatment unit with respect to the planning CTs were recorded for 30 patients with NPC. The differences in residual setup errors between XGS-10 system and CBCT were computed and quantitatively analyzed. The time of image acquisition and image registration was recorded. The radiation doses delivered by CBCT and XGS-10 were measured using PTW0.6CC ionization chambers and a water phantom. The differences between setup errors measured by the XGS-10 system and CBCT were generally <1.5 mm for translations, indicating a reasonably good agreement between the two systems for patients with NPC in the translation directions of A-P (P = 0.856), L-R (P = 0.856) and S-I (P = 0.765). Moreover, compared with CBCT, XGS-10 took much shorter image acquisition and registration time (P <0.001) and delivered only a small fraction of extra radiation dose to the patients (P <0.001). These results indicate that XGS-10 offers high localization accuracy similar to CBCT and additional benefits including prompt imaging process, low imaging radiation exposure, real time monitoring, which therefore represents a potential attractive alternative to CBCT for clinical use.  相似文献   

设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号