1.

Aim

Using flattened and unflattened photon beams, this study investigated the spectral variations of surface photon energy and energy fluence in the bone heterogeneity and beam obliquity.

Background

Surface dose enhancement is a dosimetric concern when using unflattened photon beam in radiotherapy. It is because the unflattened photon beam contains more low-energy photons which are removed by the flattening filter of the flattened photon beam.

Materials and methods

We used a water and bone heterogeneity phantom to study the distributions of energy, energy fluence and mean energy of the 6 MV flattened and unflattened photon beams (field size = 10 cm × 10 cm) produced by a Varian TrueBEAM linear accelerator. These elements were calculated at the phantom surfaces using Monte Carlo simulations. The photon energy and energy fluence calculations were repeated with the beam angle turned from 0° to 15°, 30° and 45° in the water and bone phantom.

Results

Spectral results at the phantom surfaces showed that the unflattened photon beams contained more photons concentrated mainly in the low-energy range (0–2 MeV) than the flattened beams associated with a flattening filter. With a bone layer of 1 cm under the phantom surface and within the build-up region of the 6 MV photon beam, it is found that both the flattened and unflattened beams had slightly less photons in the energy range <0.4 MeV compared to the water phantom. This shows that the presence of the bone decreased the low-energy photon backscatters to the phantom surface. When both the flattened and unflattened photon beams were rotated from 0° to 45°, the number of photon and mean photon energy increased. This indicates that both photon beams became more hardened or penetrate when the beam angle increased. In the presence of bone, the mean energies of both photon beams increased. This is due to the absorption of low-energy photons by the bone, resulting in more beam hardening.

Conclusions

This study explores the spectral relationships of surface photon energy and energy fluence with bone heterogeneity and beam obliquity for the flattened and unflattened photon beams. The photon spectral information is important in studies on the patient''s surface dose enhancement using unflattened photon beams in radiotherapy.  相似文献   

2.

Background

Medical Linear accelerators manufactured without flattening filters are increasing popular in recent days. The removal of flattening filter results in increased dose rate, reduced mean energy, reduction in head leakage and lateral scattering, which have shown advantageous when used for special treatment procedures.

Aim

This study aims to analyze physical parameters of FFF beams and to determine the inflection point for standardizing the beam flatness and penumbra.

Materials and methods

The beam profiles and depth dose patterns were measured using Radiation Field Analyzer (RFA) with 0.13 cc cylindrical ion chamber. The beam energy characteristics, head scatter factor (Sc) were obtained for 6FFF and 10FFF beams and compared with 6 MV and 10 MV photons, respectively. The symmetry and stability of unflattened regions were also analyzed. In addition, the study proposes a simple physical concept for obtaining inflection point for FFF beams and results were compared using the Akima spline interpolation method. The inflection point was used to determine the field size and penumbra of FFF beams.

Results

The Sc varied from 0.922 to 1.044 for 6FFF and from 0.913 to 1.044 for 10FFF with field sizes from 3 cm × 3 cm to 40 cm × 40 cm which is much less than FF beams. The obtained value of field size and penumbra for both simple physical concept and Akima spline interpolation methods is within the ±1.0 mm for the field size and ±2 mm penumbra. The results indicate that FFF beams reduce Sc compared with FF beams due to the absence of a flattening filter.

Conclusion

The proposed simple method to find field size and penumbra using inflection point can be accepted as it is closely approximated to mathematical results. Stability of these parameters was ascertained by repeated measurements and the study indicates good stability for FFF beam similar to that of FF beams.  相似文献   

3.
4.

Aim

The aim of the paper is to examine the relation between the increase of the photon dose in water in the region of electronic disequilibrium – so-called build-up region – and the distance of the bolus from the water surface for the applied parameters of X-ray beams.

Materials and methods

PDD measurements were carried out using the plane-parallel ionization chamber Markus in the automatic water phantom IBA BluePhantom with OmniPro-Accept V7 (IBA Dosimetry GmbH, Schwarzenbruck, Germany). All measurements were performed for different field sizes and for 6 MV and 15 MV X-ray beams, respectively. A water-equivalent RW3 slab (Goettingen White Water) produced by PTW was used as a bolus.

Results

Placing a bolus in an irradiated field changes the shape of the PDD curve in the build-up region in comparison with the one obtained for an open field. All results has been inserted in tables and figures.

Conclusion

The closer the bolus is to the water surface, the smaller the depth of the maximum dose in the phantom for all investigated fields and energies. The changes in the build-up region are important, even if the bolus does not touch the surface of the water phantom. The influence of the bolus can be ignored when the bolus-surface distance equals 25 cm for 6MV X-ray beams and 39 cm for 15 MV X-ray beams.  相似文献   

5.

Aim

The aim of the present study is to develop and verify the single film calibration procedure used in intensity-modulated radiation therapy (IMRT) quality assurance.

Background

Radiographic films have been regularly used in routine commissioning of treatment modalities and verification of treatment planning system (TPS). The radiation dosimetery based on radiographic films has ability to give absolute two-dimension dose distribution and prefer for the IMRT quality assurance. However, the single therapy verification film gives a quick and significant reliable method for IMRT verification.

Materials and methods

A single extended dose rate (EDR 2) film was used to generate the sensitometric curve of film optical density and radiation dose. EDR 2 film was exposed with nine 6 cm × 6 cm fields of 6 MV photon beam obtained from a medical linear accelerator at 5-cm depth in solid water phantom. The nine regions of single film were exposed with radiation doses raging from 10 to 362 cGy. The actual dose measurements inside the field regions were performed using 0.6 cm3 ionization chamber. The exposed film was processed after irradiation using a VIDAR film scanner and the value of optical density was noted for each region. Ten IMRT plans of head and neck carcinoma were used for verification using a dynamic IMRT technique, and evaluated using the gamma index method with TPS calculated dose distribution.

Results

Sensitometric curve has been generated using a single film exposed at nine field region to check quantitative dose verifications of IMRT treatments. The radiation scattered factor was observed to decrease exponentially with the increase in the distance from the centre of each field region. The IMRT plans based on calibration curve were verified using the gamma index method and found to be within acceptable criteria.

Conclusion

The single film method proved to be superior to the traditional calibration method and produce fast daily film calibration for highly accurate IMRT verification.  相似文献   

6.

Aim

The RapidArc commissioning and Acceptance Testing program will test and ensure accuracy in DMLC position, precise dose-rate control during gantry rotation and accurate control of gantry speed.

Background

Recently, we have upgraded our linear accelerator capable of performing IMRT which was functional from 2007 with image guided RapidArc facility. The installation of VMAT in the existing linear accelerator is a tedious process which requires many quality assurance procedures before the proper commissioning of the facility and these procedures are discussed in this study.

Materials and methods

Output of the machine at different dose rates was measured to verify its consistency at different dose rates. Monitor and chamber linearity at different dose rates were checked. DMLC QA comprising of MLC transmission factor measurement and dosimetric leaf gap measurements were performed using 0.13 cm3 and 0.65 cm3 Farmer type ionization chamber, dose 1 dosimeter, and IAEA 30 cm × 30 cm × 30 cm water phantom. Picket fence test, garden fence test, tests to check leaf positioning accuracy due to carriage movement, calibration of the leaves, leaf speed stability effects due to the acceleration and deceleration of leaves, accuracy and calibration of leaves in producing complex fields, effects of interleaf friction, etc. were verified using EDR2 therapy films, Vidar scanner, Omnipro accept software, amorphous silicon based electronic portal imaging device and EPIQA software.1–8

Results

All the DMLC related quality assurance tests were performed and evaluated by film dosimetry, portal dosimetry and EPIQA.7

Conclusion

Results confirmed that the linear accelerator is capable of performing accurate VMAT.  相似文献   

7.

Aim

The purpose of this study was to examine the usefulness of using Simultaneous Integrated Boost (SIB) radiotherapy for thyroid cancer treatment.

Background

At our hospital a 3D Conformal RadioTherapy (3D-CRT) technique involving photon and electron beams for the treatment of thyroid cancer was often used.1 High dose to the spinal canal was limiting the total dose of such a treatment. After investigation of Intensity Modulated Radiotherapy (IMRT) technique involving seven photon beams for first course of treatment3 we decided to examine possibility of reducing treatment fractions by using SIB radiotherapy.

Material and methods

Plans for 10 patients were studied. For each patient, IMRT plan for the first course of treatment (50 Gy for PTV), two plans for the second course of treatment (10 Gy for BOOST) and a SIB plan (50 Gy for PTV, 56 Gy for BOOST) were prepared. For all plans, comparisons of dose statistics for the PTV, BOOST, PTV without BOOST (defined as PTV without BOOST with 1 cm margin), spinal canal and Patient Outline (Body) was done.

Results

Minimum dose for BOOST is higher in the SIB technique than in the two course treatment. PTV without BOOST receives the same average dose in SIB and the 1st course IMRT – 50.10 Gy and 49.84 Gy, respectively. In the SIB technique, higher reduction of dose delivered to the spinal canal is possible (27 Gy compared with 30 Gy).

Conclusion

SIB therapy for thyroid cancer with relation to typical two course treatment is a good proposal of reducing the number of fractions with the same dose for BOOST and PTV without BOOST. Additionally, better sparing of the spinal canal is achieved.  相似文献   

8.

Background

Intraoperative radiotherapy (IORT) refers to the delivery of a high dose of radiation at the time of surgery.

Aim

To analyze clinical and research-oriented innovative activities developed in a 17-year period using intraoperative electron-radiation therapy (IOeRT) as a component of treatment in a multidisciplinary approach for cancer management.

Materials and methods

From 01/1995 to 03/2012 IOeRT procedures were registered in a specific Hospital-based database. Research and developments in imaging and recording for treatment planning implementation are active since 2006.

Results

1004 patients were treated and 1036 IORT procedures completed. Median age of patients was 61 (range 5 months to 94 years). Gender distribution was male in 54% of cases and female in 46%. Disease status at the time of IORT was 796 (77%) primary and 240 (23%) recurrent. Cancer type distribution included: 62% gastrointestinal, 18% sarcoma, 5% pancreas, 2% paediatric, 3% breast, 77 7% oligotopic recurrences, 2% other. IORT technical characteristics were: Applicator size 5 cm 22%, 6 cm 21%, 7 cm 21%, 8 cm 15%, 9 cm 6%, 10 cm 7% 12 cm 5% 15 cm 3%. Electron energies: 6 MeV 19%, 8 MeV 15%, 10 MeV 15%, 12 MeV 23%, 15 MeV 19%, 18 MeV 6%, other 3%. Multiple fields: 108 (11%). Dose: 7.5 Gy 3%, 10 Gy 35%, 12 Gy 3%, 12.5 Gy 49%, 15 Gy 5%, other 5%.

Conclusion

An IORT programme developed in an Academic Hospital based on practice-oriented medical decisions is an attractive interdisciplinary oncology initiative proven to be able to generate an intensive clinical activity for cancer patient quality care and a competitive source of scientific patient-oriented research, development and innovation.  相似文献   

9.

Aim

This study compared the dosimetric impact between prostate IMRT and VMAT due to patient''s weight loss.

Background

Dosimetric variation due to change of patient''s body contour is difficult to predict in prostate IMRT and VMAT, since a large number of small and irregular segmental fields is used in the delivery.

Materials and methods

Five patients with prostate volumes ranging from 32.0 to 86.5 cm3 and a heterogeneous pelvis phantom were used for prostate IMRT and VMAT plans using the same set of dose–volume constraints. Doses in IMRT and VMAT plans were recalculated with the patient''s and phantom''s body contour reduced by 0.5–2 cm to mimic size reduction. Dose coverage/criteria of the PTV and CTV and critical organs (rectum, bladder and femoral heads) were compared between IMRT and VMAT.

Results

In IMRT plans, increases of the D99% for the PTV and CTV were equal to 4.0 ± 0.1% per cm of reduced depth, which were higher than those in VMAT plans (2.7 ± 0.24% per cm). Moreover, increases of the D30% of the rectum and bladder per reduced depth in IMRT plans (4.0 ± 0.2% per cm and 3.5 ± 0.5% per cm) were higher than those of VMAT (2.2 ± 0.2% per cm and 2.0 ± 0.6% per cm). This was also true for the increase of the D5% for the right femoral head in a patient or phantom with size reduction due to weight loss.

Conclusions

VMAT would be preferred to IMRT in prostate radiotherapy, when a patient has potential to suffer from weight loss during the treatment.  相似文献   

10.

Aim

The purpose of the study was to examine the energy dependence of Gafchromic EBT radiochromic dosimetry films, in order to assess their potential use in intensity-modulated radiotherapy (IMRT) verifications.

Materials and methods

The film samples were irradiated with doses from 0.1 to 12 Gy using photon beams from the energy range 1.25 MeV to 25 MV and the film response was measured using a flat-bed scanner. The samples were scanned and the film responses for different beam energies were compared.

Results

A high uncertainty in readout of the film response was observed for samples irradiated with doses lower than 1 Gy. The relative difference exceeds 20% for doses lower than 1 Gy while for doses over 1 Gy the measured film response differs by less than 5% for the whole examined energy range. The achieved uncertainty of the experimental procedure does not reveal any energy dependence of Gafchromic EBT film response in the investigated energy range.

Conclusions

Gafchromic EBT film does not show any energy dependence in the conditions typical for IMRT but the doses measured for pre-treatment plan verifications should exceed 1 Gy.  相似文献   

11.

Aim

To study the influence of segment width on plan quality for volumetric modulated arc based stereotactic body radiotherapy.

Background

The redundancy of modulation for regularly shaped small volume tumors results in creation of many small segments and an increase of monitor units, with a consequent prolongation of treatment and uncertainty in treatment delivery.

Materials and methods

Six cases each in lung, abdomen and liver were taken for the study. For each case, three VMAT SBRT plans were generated with different penalties on minimum segment width of 0.5, 1.0 and 1.5 cm. A comparison was made on the metrics of dose volume histogram, dosimetric indices, monitor units (MUs) and delivery accuracy.

Results

The mean reduction of total MUs when compared with 0.5 cm plan was observed as 12.7 ± 6.0% and 17.5 ± 7.2% for 1.0 cm and 1.5 cm of minimum segment width, respectively. The p value showed a significant degradation in dosimetric indices for 1.5 cm plans when compared with 0.5 cm and 1.0 cm plans. The average deviation of measured dose with TPS calculated was 3.0 ± 1.1%, 2.1 ± 0.84% and 1.8 ± 0.9% for 0.5, 1.0 and 1.5 cm, respectively. The calculated gamma index with pass criteria of 2% dose difference and 2 mm distance to agreement was 95.9 ± 2.8%, 96.5 ± 2.6% and 97.8 ± 1.6% as calculated for 0.5, 1.0 and 1.5 cm of penalties, respectively. In view of the trade off between delivery efficiency and plan quality, 1 cm minimum segment width plans showed an improvement.

Conclusions

VMAT SBRT plans with increased optimal value of minimum segment width showed better plan quality and delivery efficiency for stereotactic body radiotherapy.  相似文献   

12.

Aim

The aim of this study is to evaluate differences in terms of the setup errors observed using kV planar image compared to CBCT for oesophageal cancer patients.

Background

Planar kV images are quick to acquire but only allow the observation of bony structures. CBCT allows the evaluation of soft tissues, which includes the oesophagus (and tumour) and OAR, giving a more accurate verification of the positioning.

Materials and Methods

All patients were imaged with both techniques between January 2012 and March 2014 were included in the study (16 patients, 212 kV images and 116 CBCT images). Differences between the setup errors observed on the two images modalities were studied. A correlation study between TNM staging, tumour location and immobilization systems with setup errors was also done. Finally, the calculation of systematic and random errors allowed to determine the CTV–PTV margin.

Results

A significant discrepancy (p < 0.05) between the setup errors observed with kV and CBCT was observed in the lateral direction. No statistical correlation was found between setup errors and tumour location, immobilization system or TNM staging. The CTV–PTV margin was smaller with CBCT in the vertical (0.6 cm vs. 0.9 cm) and longitudinal (0.7 cm vs. 1 cm) directions and smaller with kV for the lateral directions (0.8 cm vs. 0.9 cm).

Conclusions

The chosen modality influences the setup error observed which will influence the correction applied. Allowing a better observation of the volumes of interest, CBCT should be the modality of choice in this pathology. The CTV–PTV margins could be shrunk if CBCT is used.  相似文献   

13.

Aim

Stepping source in brachytherapy systems is used to treat a target lesion longer than the effective treatment length of the source. Cancerous lesions in the cervix, esophagus and rectum are examples of such a target lesion.

Background

In this study, the stepping source of a GZP6 afterloading intracavitary brachytherapy unit was simulated using Monte Carlo (MC) simulation and the results were used for the validation of the GZP6 treatment planning system (TPS).

Materials and methods

The stepping source was simulated using MCNPX Monte Carlo code. Dose distributions in the longitudinal plane were obtained by using a matrix shift method for esophageal tumor lengths of 8 and 10 cm. A mesh tally has been employed for the absorbed dose calculation in a cylindrical water phantom. A total of 5 × 108 photon histories were scored and the MC statistical error obtained was at the range of 0.008–3.5%, an average of 0.2%.

Results

The acquired MC and TPS isodose curves were compared and it was shown that the dose distributions in the longitudinal plane were relatively coincidental. In the transverse direction, a maximum dose difference of 7% and 5% was observed for tumor lengths of 8 and 10 cm, respectively.

Conclusion

Considering that the certified source activity is given with ±10% uncertainty, the obtained difference is reasonable. It can be concluded that the accuracy of the dose distributions produced by GZP6 TPS for the stepping source is acceptable for its clinical applications.  相似文献   

14.

Aim

Our aim was to improve dose distribution to the left breast and to determine the dose received by the ipsilateral lung, heart, contralateral lung and contralateral breast during primary left-sided breast irradiation by using intensity modulated radiotherapy (IMRT) techniques compared to conventional tangential techniques (CTT). At the same time, different beams of IMRT plans were compared to each other in respect to CI, HI and organs at risk (OAR) dose.

Background

Conventional early breast cancer treatment consists of lumpectomy followed by whole breast radiation therapy. CTT is a traditional method used for whole breast radiotherapy and includes standard wedged tangents (two opposed wedged tangential photon beams). The IMRT technique has been widely used for many treatment sites, allowing both improved sparing of normal tissues and more conformal dose distributions. IMRT is a new technique for whole breast radiotherapy. IMRT is used to improve conformity and homogeneity and used to reduce OAR doses.

Materials and methods

Thirty patients with left-sided breast carcinoma were treated between 2005 and 2008 using 6, 18 or mixed 6/18 MV photons for primary breast irradiation following breast conserving surgery (BCS). The clinical target volume [CTV] was contoured as a target volume and the contralateral breast, ipsilateral lung, contralateral lung and heart tissues as organs at risk (OAR). IMRT with seven beams (IMRT7), nine beams (IMRT9) and 11 beams (IMRT11) plans were developed and compared with CTT and among each other. The conformity index (CI), homogeneity index (HI), and doses to OAR were compared to each other.

Results

All of IMRT plans significantly improved CI (CTT: 0.76; IMRT7: 0.84; IMRT9: 0.84; IMRT11: 0.85), HI (CTT: 1.16; IMRT7: 1.12; IMRT9: 1.11; IMRT11: 1.11), volume of the ipsilateral lung receiving more than 20 Gy (>V20 Gy) (CTT: 14.6; IMRT7: 9.08; IMRT9: 8.10; IMRT11: 8.60), and volume of the heart receiving more than 30 Gy (>V30 Gy) (CTT: 6.7; IMRT7: 4.04; IMRT9: 2.80; IMRT11: 2.98) compared to CTT. All IMRT plans were found to significantly decrease >V20 Gy and >V30 Gy volumes compared to conformal plans. But IMRT plans increased the volume of OAR receiving low dose radiotherapy: volume of contralateral lung receiving 5 and 10 Gy (CTT: 0.0–0.0; IMRT7: 19.0–0.7; IMRT9: 17.2–0.66; IMRT11: 18.7–0.58, respectively) and volume of contralateral breast receiving 10 Gy (CTT: 0.03; IMRT7: 0.38; IMRT9: 0.60; IMRT11: 0.68). The differences among IMRT plans with increased number of beams were not statistically significant.

Conclusion

IMRT significantly improved conformity and homogeneity index for plans. Heart and lung volumes receiving high doses were decreased, but OAR receiving low doses was increased.  相似文献   

15.

Background

As a routine method for stepping source simulation, a Monte Carlo program is run according to the number of steps and then the summation of dose from each run is taken to obtain total dose distribution. This method is time consuming.

Aim

As an alternative method, a matrix shift based technique was applied to simulate a stepping source for brachytherapy.

Materials and methods

The stepping source of GZP6 brachytherapy unit was simulated. In a matrix shift method, it is assumed that a radiation source is stationary and instead the data matrix is shifted based on the number of steps. In this study, by running MCNPX program for one point and calculation of the dose matrix using the matrix shift method, the isodose curves for the esophageal cancer tumor lengths of 4 and 6 cm were obtained and compared with the isodose curves obtained by running MCNPX programs in each step position separately (15 and 23 steps for esophageal cancer tumor lengths of 4 and 6 cm, respectively).

Results

The difference between the two dose matrixes for the stepping and matrix shift methods based on the average dose differences are 3.85 × 10−4 Gy and 5.19 × 10−4 Gy for treatment length of 4 cm and 6 cm, respectively. Dose differences are insignificant and these two methods are equally valid.

Conclusions

The matrix shift method presented in this study can be used for calculation of dose distribution for a brachytherapy stepping source as a quicker tool compared to other routine Monte Carlo based methods.  相似文献   

16.

Aim

To evaluate the differential effects of fractionated vs. high-dose radiotherapy on plant growth.

Background

Interest in hypofractionated radiotherapy has increased substantially in recent years as tumours (especially of the lung, prostate, and liver) can be irradiated with ever greater accuracy due to technological improvements. The effects of low-dose ionizing radiation on plant growth have been studied extensively, yet few studies have investigated the effect of high-dose, hypofractionated radiotherapy on plant growth development.

Materials and methods

A total of 150 plants from the genus Capsicum annuum were randomized to receive fractionated radiotherapy (5 doses of 10 Gy each), single high-dose (SHD) radiotherapy (single 50 Gy dose), or no radiotherapy (control group). Irradiation was delivered via linear accelerator and all samples were followed daily for 26 days to assess and compare daily growth.

Results

On day 26, plants in the control, fractionated, and SHD groups had grown to a mean height of 7.55 cm, 4.32 cm, and 2.94 cm, respectively. These differences in overall growth were highly significant (P = 0.005). The SHD group showed the least amount of growth.

Conclusions

SHD effectively stunts plant growth and development. Despite the evident differences between plant and animal cells, ionizing radiation is believed to work in a similar manner in all biological cells. These findings highlight the need to continue investigating the use of hypofractionated schemes in humans to improve cancer treatment outcomes.  相似文献   

17.

Aim

To study the use of RapidArc techniques in the treatment of prostate cancer patients with hip prosthesis.

Background

An important aspect of treatment planning is to achieve dose homogeneity inside the planning target volume (PTV). Especially for those patients presenting with hip prosthesis, it becomes a challenging task to achieve dose uniformity inside the PTV.

Materials and methods

Five prostate patients presenting with hip prosthesis who had undergone radical radiotherapy were selected for this study. Depending on the composition of prosthesis, a predefined set of Hounsfield values were assigned to each study set. RapidArc plans were generated on an Eclipse treatment planning system. Two arcs that include clockwise and counter-clockwise arcs were used in all these cases. To avoid beams passing through the prosthesis, a simple structure was defined around it with 1 cm margin and a strict dose constraint applied to the block during VMAT optimization.

Results

The mean D2/D98 ratio of PTV for all the patients was 1.06 ± 0.01. The mean percentage rectum volume receiving 50 Gy, 60 Gy, 70 Gy and 75 Gy for all the patients were 33.1 ± 5.9, 21.7 ± 5.5, 13.8 ± 4.4 and 9.5 ± 3.0, respectively.

Conclusions

This study shows that using a double arc RapidArc technique is a simple and effective treatment method of treating prostate cancer in patients presenting with a hip prosthesis. The definition of a beam avoidance structure encompassing the prosthesis and applying strict dose constraints to it reduces the beam contribution to the prosthesis  相似文献   

18.

Aim

To employ the thermal neutron background that affects the patient during a traditional high-energy radiotherapy treatment for BNCT (Boron Neutron Capture Therapy) in order to enhance radiotherapy effectiveness.

Background

Conventional high-energy (15–25 MV) linear accelerators (LINACs) for radiotherapy produce fast secondary neutrons in the gantry with a mean energy of about 1 MeV due to (γ, n) reaction. This neutron flux, isotropically distributed, is considered as an unavoidable undesired dose during the treatment. Considering the moderating effect of human body, a thermal neutron fluence is localized in the tumour area: this neutron background could be employed for BNCT by previously administering 10B-Phenyl-Alanine (10BPA) to the patient.

Materials and methods

Monte Carlo simulations (MCNP4B-GN code) were performed to estimate the total amount of neutrons outside and inside human body during a traditional X-ray radiotherapy treatment.Moreover, a simplified tissue equivalent anthropomorphic phantom was used together with bubble detectors for thermal and fast neutron to evaluate the moderation effect of human body.

Results

Simulation and experimental results confirm the thermal neutron background during radiotherapy of 1.55E07 cm−2 Gy−1.The BNCT equivalent dose delivered at 4 cm depth in phantom is 1.5 mGy-eq/Gy, that is about 3 Gy-eq (4% of X-rays dose) for a 70 Gy IMRT treatment.

Conclusions

The thermal neutron component during a traditional high-energy radiotherapy treatment could produce a localized BNCT effect, with a localized therapeutic dose enhancement, corresponding to 4% or more of photon dose, following tumour characteristics. This BNCT additional dose could thus improve radiotherapy, acting as a localized radio-sensitizer.  相似文献   

19.

Background

The most often found complications in patients with breast cancer who received radiotherapy are cardiac and pulmonary function disorders and development of second malignancies.

Aim

To compare the intensity modulated radiotherapy with the 3D tangential beams technique in respect of dose distribution in target volume and critical organs they generate in patients with early-stage breast cancer who received breast-conserving therapy.

Materials and methods

A dosimetric analysis was performed to assess the three radiotherapy techniques used in each of 10 consecutive patients with early-stage breast cancer treated with breast-conserving therapy. Radiotherapy was planned with the use of all the three techniques: 3D tangential beams with electron boost, IMRT with electron boost, and intensity modulated radiotherapy with simultaneous integrated boost.

Results

The use of the IMRT techniques enables more homogenous dose distribution in target volume. The range of mean and median dose to the heart and lung was lower with the IMRT techniques in comparison to the 3D tangential beams technique. The range of mean dose to the heart amounted to 0.3–3.5 Gy for the IMRT techniques and 0.4–4.3 for the tangential beams technique. The median dose to the lung on the irradiated side amounted to 4.9–5 Gy for the IMRT techniques and 5.6 Gy for the 3D tangential beams technique.

Conclusion

The application of the IMRT techniques in radiotherapy patients with early-stage breast cancer allows to obtain more homogenous dose distribution in target volume, while permitting to reduce the dose to critical organs.  相似文献   

20.

Background

Low-dose rate brachytherapy is a well established treatment modality of oral cancer. Data about high-dose rate (HDR) brachytherapy are still sparse with various fractionation schedules and heterogeneous results.

Aim

The aim of our retrospective study was to evaluate the results of HDR brachytherapy with doses of 3 Gy twice daily.

Patients and methods

Twenty patients with squamous cell tongue cancer were treated in the years 2001–2009 by exclusive HDR BT 18 × 3 Gy twice daily. The plastic tube technique was used. Median follow up was 47 months (7.8–118) since brachytherapy.

Results

The local and locoregional control was 85% and 68%, respectively. Bone necrosis developed in one case treated without mandibular shielding and soft tissue necrosis in 2 cases.

Conclusion

It can be concluded that HDR brachytherapy with 18 × 3 Gy twice daily is safe with promising local control. The risk of nodal recurrences is substantial.  相似文献   