1.

Aim

Using flattened and unflattened photon beams, this study investigated the spectral variations of surface photon energy and energy fluence in the bone heterogeneity and beam obliquity.

Background

Surface dose enhancement is a dosimetric concern when using unflattened photon beam in radiotherapy. It is because the unflattened photon beam contains more low-energy photons which are removed by the flattening filter of the flattened photon beam.

Materials and methods

We used a water and bone heterogeneity phantom to study the distributions of energy, energy fluence and mean energy of the 6 MV flattened and unflattened photon beams (field size = 10 cm × 10 cm) produced by a Varian TrueBEAM linear accelerator. These elements were calculated at the phantom surfaces using Monte Carlo simulations. The photon energy and energy fluence calculations were repeated with the beam angle turned from 0° to 15°, 30° and 45° in the water and bone phantom.

Results

Spectral results at the phantom surfaces showed that the unflattened photon beams contained more photons concentrated mainly in the low-energy range (0–2 MeV) than the flattened beams associated with a flattening filter. With a bone layer of 1 cm under the phantom surface and within the build-up region of the 6 MV photon beam, it is found that both the flattened and unflattened beams had slightly less photons in the energy range <0.4 MeV compared to the water phantom. This shows that the presence of the bone decreased the low-energy photon backscatters to the phantom surface. When both the flattened and unflattened photon beams were rotated from 0° to 45°, the number of photon and mean photon energy increased. This indicates that both photon beams became more hardened or penetrate when the beam angle increased. In the presence of bone, the mean energies of both photon beams increased. This is due to the absorption of low-energy photons by the bone, resulting in more beam hardening.

Conclusions

This study explores the spectral relationships of surface photon energy and energy fluence with bone heterogeneity and beam obliquity for the flattened and unflattened photon beams. The photon spectral information is important in studies on the patient''s surface dose enhancement using unflattened photon beams in radiotherapy.  相似文献   

2.
PurposeOver the last decades, Gold Nanoparticles (AuNPs) have been presented as an innovative approach in radiotherapy (RT) enhancement. Several studies have proven that the irradiation of tumors containing AuNPs could lead to more effective tumor control than irradiation alone. Studies with low kV photons and AuNPs conclude in encouraging results regarding the level of radioenhancement. However, experimental and theoretical studies with MV photons report controversial findings concerning the correlation between dose enhancement effect and tumor cell killing. The great variation in the experimental protocols and simulations complicates the comparison of their outcomes and depicts the need for limiting the variety of investigated parameters. Our purpose is to point out a possible direction for building realistic Monte Carlo (MC) models that could end up with promising results in MV photons RT enhancement.MethodsWe explored published in silico studies concerning AuNPs enhanced RT from 2010 to 2019. In this review, we discuss the different AuNPs and MV photon beams characteristics that have been reported and their effect in dose enhancement.ResultsAuNPs size, concentration, type of distribution along with photon beams energy and the presence of flattening filter in linear accelerators seem to be the major parameters that determine AuNPs radioenhancement in silico.ConclusionsPrior to AuNPs clinical translation in photon radiotherapy, in silico studies should emphasize on nanodosimetry and track structure codes than condensed history ones. Toxicity estimation and biological aspects should be implemented in MC simulations so as to achieve accurate and realistic modelling of AuNPs driven RT.  相似文献   

3.
BackgroundHigh-energy photon and electron therapeutic beams generated in medical linear accelerators can cause the electronuclear and photonuclear reactions in which neutrons with a broad energy spectrum are produced. A low-energy component of this neutron radiation induces simple capture reactions from which various radioisotopes originate and in which the radioactivity of a linac head and various objects in the treatment room appear.AimThe aim of this paper is to present the results of the thermal/resonance neutron fluence measurements during therapeutic beam emission and exemplary spectra of gamma radiation emitted by medical linac components activated in neutron reactions for four X-ray beams and for four electron beams generated by various manufacturers’ accelerators installed in typical concrete bunkers in Polish oncological centers.Materials and methodsThe measurements of neutron fluence were performed with the use of the induced activity method, whereas the spectra of gamma radiation from decays of the resulting radioisotopes were measured by means of a portable high-purity germanium detector set for field spectroscopy.ResultsThe fluence of thermal neutrons as well as resonance neutrons connected with the emission of a 20 MV X-ray beam is ~106 neutrons/cm2 per 1 Gy of a dose in water at a reference depth. It is about one order of magnitude greater than that for the 15 MV X-ray beams and about two orders of magnitude greater than for the 18–22 MeV electron beams regardless of the type of an accelerator.ConclusionThe thermal as well as resonance neutron fluence depends strongly on the type and the nominal potential of a therapeutic beam. It is greater for X-ray beams than for electrons. The accelerator accessories and other large objects should not be stored in a treatment room during high-energy therapeutic beam emission to avoid their activation caused by thermal and resonance neutrons. Half-lives of the radioisotopes originating from the simple capture reaction (n,γ) (from minutes to hours) are long enough to accumulate radioactivity of components of the accelerator head. The radiation emitted by induced radioisotopes causes the additional doses to staff operating the accelerators.  相似文献   

4.
AimThe aim of this study was to characterize the radiation contamination inside and outside the megavoltage radiotherapy room.BackgroundRadiation contamination components in the 18 MV linac room are the secondary neutron, prompt gamma ray, electron and linac leakage radiation.Materials and MethodsAn 18 MV linac modeled in a typical bunker employing the MCNPX code of Monte Carlo. For fast calculation, phase-space distribution (PSD) file modeling was applied and the calculations were conducted for the radiation contamination components dose and spectra at 6 locations inside and outside the bunker.ResultsThe results showed that the difference of measured and calculated percent depth-dose (PDD) and photo beam-profile (PBP) datasets were lower than acceptable values. At isocenter, the obtained photon dose and neutron fluence were 2.4 × 10−14 Gy/initial e° and 2.22 × 10-8 n°/cm2, respectively. Then, neutron apparent source strength (QN) value was found as 1.34 × 1012 n°/Gy X at isocenter and the model verified to photon and neutron calculations. A surface at 2 cm below the flattening filter was modeled as phase-space (PS) file for PDD and PBP calculations. Then by use of a spherical cell in the center of the linac target as a PS surface, contaminant radiations dose, fluence and spectra were estimated at 6 locations in a considerably short time, using the registered history of all particles and photons in the 13GB PSD file as primary source in the second step.ConclusionDesigning the PSD file in MC modeling helps user to solve the problems with complex geometry and physics precisely in a shorter run-time.  相似文献   

5.
AimThis study aimed to commission the Elekta Infinity™ working in 6 and 10 MV photon beam installed in Concord International Hospital, Singapore, and compare the OFs between MC simulation and measurement using PTW semiflex and microDiamond detector for small field sizes.Material and MethodsThere are two main steps in this study: modelling of Linac 6 and 10 MV photon beam and analysis of the output factors for field size 2 × 2–10 × 10 cm2. The EGSnrc/BEAMnrc-DOSXYZnrc code was used to model and characterize the Linac and to calculate the dose distributions in a water phantom. The dose distribution and OFs were compared to the measurement data in the same condition.ResultsThe commissioning process was only conducted for a 10 × 10 cm2 field size. The PDD obtained from MC simulation showed a good agreement with the measurement. The local dose difference of PDDs was less than 2% for 6 and 10 MV. The initial electron energy was 5.2 and 9.4 MeV for 6 and 10 MV photon beam, respectively. This Linac model can be used for dose calculation in other situations and different field sizes because this Linac has been commissioned and validated using Monte Carlo simulation. The 10 MV Linac produces higher electron contamination than that of 6 MV.ConclusionsThe Linac model in this study was acceptable. The most important result in this work comes from OFs resulted from MC calculation. This value was more significant than the OFs from measurement using semiflex and microDiamond for all beam energy and field sizes because of the CPE phenomenon.  相似文献   

6.
This study investigates the superficial dose from FFF beams in comparison with the conventional flattened ones using a Monte Carlo (MC) method. Published phase-space files which incorporated real geometry of a TrueBeam accelerator were used for the dose calculation in phantom and clinical cases. The photon fluence on the central axis is 3 times that of a flattened beam for a 6 MV FFF beam and 5 times for a 10 MV beam. The mean energy across the field in air at the phantom surface is 0.92–0.95 MeV for the 6 MV FFF beam and 1.18–1.30 MeV for the corresponding flattened beam. At 10 MV, the values are 1.52–1.72 and 2.15–2.87 MeV for the FFF and flattened beams, respectively. The phantom dose at the depth of 1 mm in the 6 MV FFF beam is 6% ± 2.5% (of the maximum dose) higher compared to the flattened beam for a 25 × 25 cm2 field and 14.6% ± 1.9% for the 2 × 2 cm2 field. For the 10 MV beam, the corresponding differences are 3.4% ± 1.5% and 10.7% ± 0.6%. The skin dose difference at selected points on the patient's surface between the plans using FFF and flattened beams in the head-and-neck case was 6.5% ± 2.3% (1SD), and for the breast case it was 6.4% ± 2.3%. The Monte Carlo simulations showed that due to the lower mean energy in the FFF beam, the clinical superficial dose is higher without the flattening filter compared to the flattened beam.  相似文献   

7.
PurposeWe investigated the impact of transverse magnetic fields on the dose response of a nanoDot optically stimulated luminescence dosimetry (OSLD) in megavoltage photon beams.MethodsThe nanoDot OSLD response was calculated via Monte Carlo (MC) simulations. The responses RQ and RQ,B without and with the transverse magnetic fields of 0.35–3 T were analyzed as a function of depth at a 10 cm × 10 cm field for 4–18 MV photons in a solid water phantom. All responses were determined based on comparisons with the response under the reference conditions (depth of 10 cm and a 10 cm × 10 cm field) for 6 MV without the magnetic field. In addition, the influence of air-gaps on the nanoDot response in the magnetic field was estimated according to Burlin’s general cavity theory.ResultsThe RQ as a function of depth for 4–18 MV ranged from 1.013 to 0.993, excepting the buildup region. The RQ,B increased from 2.8% to 1.5% at 1.5 T and decreased from 3.0% to 1.1% at 3 T in comparison with RQ as the photon energy increased. The depth dependence of RQ,B was less than 1%, excepting the buildup region. The top air-gap and the bottom air- gap were responsible for the response reduction and the response increase, respectively.ConclusionsThe response RQ,B varied depending on the magnetic field intensity, and the variation of RQ,B reduced as the photon beam energy increased. The air-gaps affected the dose deposition in the magnetic fields.  相似文献   

8.
AimEvaluation of energy deposition of protons in human brain and calculation of the secondary neutrons and photons produced by protons in proton therapy.BackgroundRadiation therapy is one of the main methods of treating localized cancer tumors. The use of high energy proton beam in radiotherapy was proposed almost 60 years ago. In recent years, there has been a revival of interest in this subject in the context of radiation therapy. High energy protons suffer little angular deflection and have a well-defined penetration range, with a sharp increase in the energy loss at the end of their trajectories, namely the Bragg peak.Materials and methodsA slab head phantom was used for the purpose of simulating proton therapy in brain tissue. In this study simulation was carried out using the Monte Carlo MCNPX code.ResultsBy using mono energetic proton pencil beams, energy depositions in tissues, especially inside the brain, as well as estimating the neutron and photon production as a result of proton interactions in the body, together with their energy spectra, were calculated or obtained. The amount of energy escaped from the head by secondary neutrons and photons was determined.ConclusionsIt was found that for high energy proton beams the amount of escaped energy by neutrons is almost 10 times larger than that by photons. We estimated that at 110 MeV beam energy, the overall proton energy “leaked” from the head by secondary photons and neutrons to be around 1%.  相似文献   

9.
PurposeThis study aims to investigate the relationship between backscatter factors and Al-half-value-layers (Al-HVL) by making the quality index (QI) a parameter for diagnostic kilovoltage x-ray beams.MethodsBackscatter factors, Bw, for x-ray fluence spectra were calculated from the weighted average of Bw for monoenergetic photons of between 8 and 140 keV with field sizes of 10 cm × 10 cm to 40 cm × 40 cm. The value of Bw for monoenergetic photons was calculated from the ratio of the water kerma at the surface of a water phantom and that at the same point free-in-air using the EGSnrc/cavity code. The weighted averaged backscatter factors were validated by comparing them with those of direct Monte Carlo calculations for the x-ray fluence spectra. The Bw for the x-ray fluence spectra were classified by a QI of 0.35, 0.4, 0.5, 0.6, and 0.7 specified by the ratio of the effective energy and maximum energy. The relationship between Bw and Al-HVL was evaluated for the given QI values. The x-ray fluence spectra were generated for tube voltages of 40–140 kVp with Al-HVLs of 0.5–13.2 mm using the SpekCalc program.ResultsThe weighted averaged backscatter factors for x-ray fluence spectra agreed within 0.7% with those of the direct Monte Carlo calculations. The backscatter factors were represented by the fitting curves of R2 > 0.99 with Al-HVL for the given QI values.ConclusionsIt is possible to obtain Bw more accurately by using QI specified by the measured Al-HVL.  相似文献   

10.
AimThe thermal neutron, epithermal neutron and high-energy photon fluence were measured in this work around the Varian 21EX 23 MV CLINAC, which is operated in Albairouni hospital in Damascus, Syria.BackgroundPhotoneutron measurements around medical CLINAC aim to protect both patients and staff from unwanted radiation.Materials and methodsNeutron and photon activation techniques were applied using gold foils.ResultsIt was found that high-energy photons fluence has practically a constant value in the field size. The thermal and epithermal neutron fluence along ox and oy axes has the same order of magnitude.ConclusionGold foils have been used successfully to measure neutron flux and high-energy photons simultaneously using activation techniques.  相似文献   

11.
ObjectiveTo determine the optimum energy and beam arrangement for prostate intensity-modulated radiation therapy (IMRT) delivery using an Elekta Beam Modulator? linear accelerator, in order to inform decisions when commissioning IMRT for prostate cancer.MethodsCMS XiO was used to create IMRT plans for a prostate patient. Arrangements with 3, 5, 7, 9 and 11 equally spaced fields, containing both a direct anterior and a direct posterior beam were used, with both 6 MV and 10 MV photons. The effects of varying the maximum number of iterations, leaf increment, number of intensity levels and minimum segment size were investigated. Treatment plans were compared using isodose distributions, conformity indices for targets and critical structures, target dose homogeneity, body dose and plan complexity.ResultsTarget dose conformity and homogeneity and sparing of critical structures improved with an increasing number of beams, although any improvements were small for plans containing more than five fields. Set-ups containing a direct posterior field provided superior conformality around the rectum to anterior beam arrangements. Mean non-target dose and total number of monitor units were higher with 6 MV for all beam arrangements. The dose distribution resulting from seven 6 MV beams was considered clinically equivalent to that with five 10 MV beams.ConclusionMethods have been developed to plan IMRT treatments using XiO for delivery with a Beam Modulator? that fulfil demanding dose criteria, using many different set-ups. This study suggests that 6 MV photons can produce prostate IMRT plans that are comparable to those using 10 MV. Work is ongoing to develop a complete class solution.  相似文献   

12.
AimTo determine the energy and dose dependence of GafChromic EBT3-V3 film over an energy range 0.2 mm Al HVL to 6 MV.BackgroundThe decay scheme of a brachytherapy source may be complex and the spectrum of energy can be wide. LiF TLDs are the golden standard recommended for dosimetric measures in brachytherapy, for their energy independence, but TLDs could be not available in some centres. An alternative way to perform dose measurements is to use GafChromic films, but they show energy dependence.Methods and materialsFilms have been irradiated at increasing dose with three different beams: 6 MV beam, TPR20, 10 = (0.684 ± 0.01), HVL = (2.00 ± 0.01)mmAl and HVL = (0.20 ± 0.01)mmAl. Calibration curves were generated using the same dose range (0cGy to 850cGy) for the three energies. Using the 6 MV calibration curve as reference, the film response in terms of net optical density (OD) was evaluated.ResultsThe difference in the calibration curve obtained by irradiating the film with 6 MV and 2 mm Al HVL energy beams is less than 3 %, within the calibration uncertainty, in the dose range 500-850cGy. The OD of EBT3-V3 film is significantly lower at 0.2 mmAl HVL compared to 6 MV, showing differences up to 25 %.ConclusionWithin the range 6 MV-2 mm Al HVL and dose higher than 500cGy, GafChromic EBT3-V3 films are energy independent. In this dose range, films can be calibrated in a simple geometry, using a 6 MV Linac beam, and can be used for brachytherapy sources dose measures. The use of EBT3 films can be extended to reference dosimetry in Ir-192 clinical brachytherapy.  相似文献   

13.
Purposeto elucidate the effects of multiple scattering and energy-loss straggling on electron beams slowing down in materials.MethodsEGSnrc Monte Carlo simulations are done using a purpose-written user-code.ResultsPlots are presented of the primary electron’s energy as a function of pathlength for 20 MeV electrons incident on water and tantalum as are plots of the overall distribution of pathlengths as the 20 MeV electrons slow down under various Monte Carlo scenarios in water and tantalum. The distributions range from 1 % to 135 % of the CSDA range in water and from 1 % to 186 % in tantalum. The effects of energy-loss straggling on energy spectra at depth and electron fluence at depth are also presented.ConclusionsThe role of energy-loss straggling and multiple scattering are shown to play a significant role in the range straggling which determines the dose fall-off region in electron beam dose vs depth curves and a significant role in the energy distributions as a function of depth.  相似文献   

14.
PurposeEquipment refurbishment was performed to remove the beam-hardening filter (BHF) from the CyberKnife system (CK). This study aimed to confirm the change in the beam characteristics between the conventional CK (present-BHF CK) and CK after the BHF was removed (absent-BHF CK) and evaluate the impact of BHF removal on the beam quality correction factors kQ.MethodsThe experimental measurements of the beam characteristics of the present- and absent-BHF CKs were compared. The CKs were modeled using Monte Carlo simulations (MCs). The energy fluence spectra were calculated using MCs. Finally, kQ were estimated by combining the MC results and analytic calculations based on the TRS-398 and TRS-483 approaches.ResultsAll gamma values for percent depth doses and beam profiles between each CK were less than 0.5 following the 3%/1 mm criteria. The percentage differences for tissue-phantom ratios at depths of 20 and 10 cm and percentage depth doses at 10 cm between each CK were −1.20% and −0.97%, respectively. The MC results demonstrated that the photon energy fluence spectrum of the absent-BHF CK was softer than that of the present-BHF CK. The kQ values for the absent-BHF CK were in agreement within 0.02% with those for the present-BHF CK.ConclusionsThe photon energy fluence spectrum was softened by the removal of BHF. However, no remarkable impact was observed for the measured beam characteristics and kQ. Therefore, the previous findings of the kQ values for the present-BHF CK can be directly used for the absent-BHF CK.  相似文献   

15.
PurposeA Geant4 model of a novel, water-equivalent electronic portal imaging device (EPID) prototype for radiotherapy imaging and dosimetry utilising an array of plastic scintillating fibres (PSFs) has been developed. Monte Carlo (MC) simulations were performed to quantify the PSF-EPID imaging performance and to investigate design aspects affecting performance for optimisation.MethodsUsing the Geant4 model, the PSF-EPID’s imaging performance for 6 MV photon beams was quantified in terms of its modulation transfer function (MTF), noise power spectrum (NPS) and detective quantum efficiency (DQE). Model parameters, including fibre dimensions, optical cladding reflectivity and scintillation yield, were varied to investigate impact on imaging performance.ResultsThe MC-calculated DQE(0) for the reference PSF-EPID geometry employing 30 mm fibres was approximately nine times greater than values reported for commercial EPIDs. When using 10 mm long fibres, the PSF-EPID DQE(0) was still approximately three times greater than that of a commercial EPID. Increased fibre length, cladding reflectivity and scintillation yield produced the greatest decreases in NPS and increases in DQE.ConclusionsThe potential to develop an optimised next-generation water-equivalent EPID with MV imaging performance at least comparable to commercial EPIDs has been demonstrated. Factors most important for optimising prototype design include fibre length, cladding reflectivity and scintillation yield.  相似文献   

16.
AimIn this study, we investigated initial electron parameters of Siemens Artiste Linac with 6 MV photon beam using the Monte Carlo method.BackgroundIt is essential to define all the characteristics of initial electrons hitting the target, i.e. mean energy and full width of half maximum (FWHM) of the spatial distribution intensity, which is needed to run Monte Carlo simulations. The Monte Carlo is the most accurate method for simulation of radiotherapy treatments.Materials and methodsLinac head geometry was modeled using the BEAMnrc code. The phase space files were used as input file to DOSXYZnrc simulation to determine the dose distribution in a water phantom. We obtained percent depth dose curves and the lateral dose profile. All the results were obtained at 100 cm of SSD and for a 10 × 10 cm2 field.ResultsWe concluded that there existed a good conformity between Monte Carlo simulation and measurement data when we used electron mean energy of 6.3 MeV and 0.30 cm FWHM value as initial parameters. We observed that FWHM values had very little effect on PDD and we found that the electron mean energy and FWHM values affected the lateral dose profile. However, these effects are between tolerance values.ConclusionsThe initial parameters especially depend on components of a linac head. The phase space file which was obtained from Monte Carlo Simulation for a linac can be used as calculation of scattering, MLC leakage, to compare dose distribution on patients and in various studies.  相似文献   

17.
The aim of this study was to determine the surface doses using GafChromic EBT films and compare them with plane-parallel ionization chamber measurements for 6 and 18 MV high energy photon beams. The measurements were made in a water equivalent solid phantom in the build-up region of the 6 and 18 MV photon beams at 100 cm SSD for various field sizes. Markus type plane-parallel ion chamber with fixed-separation between collecting electrodes was used to measure the percent depth doses. GafChromic EBT film measurements were performed both on the phantom surface and maximum dose depth at the same geometry with ion chamber measurements. The surface doses found using GafChromic EBT film were 15%, 20%, 29%and 39% ± 2% (1SD) for 6 MV photons, 6%, 11%, 23% and 32% ± 2% (1SD) for 18 MV photons at 5, 10, 20 and 30 cm2 field sizes, respectively. GafChromic EBT film provides precise measurements for surface dose in the high energy photons. Agreement between film and plane-parallel chamber measurements was found to be within ±3% for 18 MV photon beams. There was 5% overestimate on the surface doses when compared with the plane-parallel chamber measurements for all field sizes in the 6 MV photon beams.  相似文献   

18.
PurposeTo evaluate the uncertainties and characteristics of radiochromic film-based dosimetry system using the EBT3 model Gafchromic® film in therapy photon, electron and proton beams.Material and methodsEBT3 films were read using an EPSON Expression 10000XL/PRO scanner. They were irradiated in five beams, an Elekta SL25 6 MV and 18 MV photon beam, an IBA 100 MeV 5 × 5 cm2 proton beam delivered by pencil-beam scanning, a 60 MeV fixed proton beam and an Elekta SL25 6 MeV electron beam. Reference dosimetry was performed using a FC65-G chamber (Elekta beam), a PPC05 (IBA beam) and both Markus 1916 and PPC40 Roos ion-chambers (60 MeV proton beam). Calibration curves of the radiochromic film dosimetry system were acquired and compared within a dose range of 0.4–10 Gy. An uncertainty budget was estimated on films irradiated by Elekta SL25 by measuring intra-film and inter-film reproducibility and uniformity; scanner uniformity and reproducibility; room light and film reading delay influences.ResultsThe global uncertainty on acquired optical densities was within 0.55% and could be reduced to 0.1% by placing films consistently at the center of the scanner. For all beam types, the calibration curves are within uncertainties of measured dose and optical densities. The total uncertainties on calibration curve due to film reading and fitting were within 1.5% for photon and proton beams. For electrons, the uncertainty was within 2% for dose superior to 0.8 Gy.ConclusionsThe low combined uncertainty observed and low beam and energy-dependence make EBT3 suitable for dosimetry in various applications.  相似文献   

19.
PurposeIn radiotherapy, accurate calculation of patient radiation dose is very important for good clinical outcome. In the presence of metallic implants, the dose calculation accuracy could be compromised by metal artefacts generated in computed tomography (CT) images of patients. This study investigates the influence of metal-induced CT artefacts on MC dose calculations in a pelvic prosthesis phantom.MethodsA pelvic phantom containing unilateral Ti prosthesis was CT-scanned and accurate Hounsfield unit (HU) values were assigned to known materials of the phantom as opposed to HU values produced through the artefact CT images of the phantom. Using the DOSXYZnrc MC code, dose calculations were computed in the phantom model constructed from the original CT images containing the artefacts and artefact-free images made from the exact geometry of the phantom with known materials. The dose calculations were benchmarked against Gafchromic EBT3 film measurements using 15 MeV electron and 10 MV photon beams.ResultsThe average deviations between film and MC dose data decreased from 3 ± 2% to 1 ± 1% and from about 6 ± 2% to 3 ± 1% for the artefact and artefact-free phantom models against film data for the electron and photon fields, respectively.ConclusionsFor the Ti prosthesis phantom, the presence of metal-induced CT artefacts could cause dose inaccuracies of about 3%. Construction of an artefact-free phantom model made from the exact geometry of the phantom with known materials to overcome the effect of artefacts is advantageous compared to using CT data directly of which the exact tissue composition is not well-known.  相似文献   

20.

Aim

The aim of this work was to map the characteristics of (n,γ) and (γ,n) reactions in a high energy photon radiation therapy.

Background

Photoneutrons produced in the high energy X-Ray radiation therapy may damage patients and staff. It is due to high RBE of the produced neutrons according to their energy and isotropic emission. Characterization of the photoneutrons can help us in appropriate shielding.

Materials and methods

This study focused on the photoneutron and capture gamma ray phenomena. Characteristics such as dose value, fluence and spectra of both the neutrons and the by produced prompt gamma ray were described.

Results and discussion

Neutron and prompt gamma spectra in different points showed the neutrons to be thermalized when increasing the distance from the linac. Energy of the neutrons changed from about 0.6 MeV at the isocentre to around 10−08 MeV at the outer door position. Although the neutrons were found as fast neutrons, their spectra showed they were thermal neutrons at the outer door position. Additionally, it was seen that the energy of the gamma rays is higher than the scattered X-ray energy. The energy of gamma rays was seen to be up to 10 MeV while the linac photons had energy lower than 1 MeV. Neutron source strength obtained in this work was in good agreement with the published data, which may be a confirmation of our simulation accuracy.

Conclusion

The study showed that the Monte Carlo simulation can be applied in the radiotherapy and industrial radiation works as a useful and precise estimator. We also concluded that the dose from the prompt gamma ray at the outer door location is higher than the scattered radiation from the linac and should be considered in the shielding.  相似文献   