1.
Krzysztof Chełmiński Wojciech Bulski Dietmar Georg Dominika Bodzak Zbigniew Maniakowski Dominika Oborska Joanna Rostkowska Małgorzata Kania 《Reports of Practical Oncology and Radiotherapy》2010,15(2):40-46
Aim
The purpose of the study was to examine the energy dependence of Gafchromic EBT radiochromic dosimetry films, in order to assess their potential use in intensity-modulated radiotherapy (IMRT) verifications.Materials and methods
The film samples were irradiated with doses from 0.1 to 12 Gy using photon beams from the energy range 1.25 MeV to 25 MV and the film response was measured using a flat-bed scanner. The samples were scanned and the film responses for different beam energies were compared.Results
A high uncertainty in readout of the film response was observed for samples irradiated with doses lower than 1 Gy. The relative difference exceeds 20% for doses lower than 1 Gy while for doses over 1 Gy the measured film response differs by less than 5% for the whole examined energy range. The achieved uncertainty of the experimental procedure does not reveal any energy dependence of Gafchromic EBT film response in the investigated energy range.Conclusions
Gafchromic EBT film does not show any energy dependence in the conditions typical for IMRT but the doses measured for pre-treatment plan verifications should exceed 1 Gy. 相似文献2.
《Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)》2016,32(10):1263-1270
PurposeAim of this work is the assessment of build-up and superficial doses of different clinical Head&Neck plans delivered with Helical TomoTherapy (HT) (Accuray, Sunnyvale, CA). Depth dose profiles and superficial dose points were measured in order to evaluate the Treatment Planning System (TPS) capability of an accurate dose modeling in regions of disequilibrium. Geometries and scattering conditions were investigated, similar to the ones generally encountered in clinical treatments.MethodsMeasurements were performed with two dosimeters: Gafchromic® EBT3 films (Ashland Inc., Wayne, NJ) and a synthetic single crystal diamond detector (PTW-Frieburg microDiamond, MD). A modified version of the Alderson RANDO phantom was employed to house the detectors. A comparison with TPS data was carried out in terms of dose difference (DD) and distance-to-agreement (DTA).ResultsDD between calculated data and MD measurements are within 4% even in points with high spatial dose variation. For depth profiles, EBT3 data show a DDmax of 3.3% and DTAmax of 2.2 mm, in low and high gradient regions, respectively, and compare well with MD data. EBT3 superficial points always results in measured doses lower than TPS evaluated ones, with a maximum DTA value of 1.5 mm.ConclusionsDoses measured with the two devices are in good agreement and compare well with calculated data. The deviations found in the present work are within the reference tolerance level, suggesting that the HT TPS is capable of a precise dose estimation both in superficial regions and in correspondence with interfaces between air and PMMA. 相似文献
3.
PurposeIntra-Operative Electron Radiation Therapy (IOERT) is used to treat rectal cancer at our institution, and in vivo measurements with Gafchromic EBT3® films were introduced as quality assurance. The purpose of this work was to quantify the uncertainties associated with digitization of very small EBT3 films irradiated simultaneously, in order to optimize in vivo dosimetry for IOERT.MethodsFilm samples of different sizes - M1 (5×5cm2), M2 (1.5×1.5 cm2), M3 (1.0×1.5 cm2) and M4 (0.75×1.5 cm2) – were used to quantify typical variations (uncertainties) due to scanner fluctuations, misalignment, film inhomogeneity, long-term effect of film cutting, small rotations, film curling, edge effects and the influence of opaque templates. Fitting functions and temporal validity of sensitometric curves were also assessed.ResultsFilm curling, intra-film variability and scanner fluctuations are important effects that need to be minimized or considered in the uncertainty budget. Small rotations, misalignments and film cutting have little or no influence on the readings. Most fitting functions perform well, but the quantity used for dose quantification determines over- or under-valuation of dose in the long term. Edge effects and the influence of opaque templates need to be well understood, to allow optimization of methodology to the intended purpose.ConclusionThe proposed method allows practical and simultaneous digitization of up to ten small irradiated film samples, with an experimental uncertainty of 1%. 相似文献
4.
PurposeTo investigate the feasibility of a fast protocol for radiochromic film dosimetry to verify intensity-modulated radiotherapy (IMRT) plans.Method and materialsEBT3 film dosimetry was conducted in this study using the triple-channel method implemented in the cloud computing application (Radiochromic.com). We described a fast protocol for radiochromic film dosimetry to obtain measurement results within 1 h.Ten IMRT plans were delivered to evaluate the feasibility of the fast protocol. The dose distribution of the verification film was derived at 15, 30, 45 min using the fast protocol and also at 24 h after completing the irradiation. The four dose maps obtained per plan were compared using global and local gamma index (5%/3 mm) with the calculated one by the treatment planning system. Gamma passing rates obtained for 15, 30 and 45 min post-exposure were compared with those obtained after 24 h.ResultsSmall differences respect to the 24 h protocol were found in the gamma passing rates obtained for films digitized at 15 min (global: 99.6% ± 0.9% vs. 99.7% ± 0.5%; local: 96.3% ± 3.4% vs. 96.3% ± 3.8%), at 30 min (global: 99.5% ± 0.9% vs. 99.7% ± 0.5%; local: 96.5% ± 3.2% vs. 96.3 ± 3.8%) and at 45 min (global: 99.2% ± 1.5% vs. 99.7% ± 0.5%; local: 96.1% ± 3.8% vs. 96.3 ± 3.8%).ConclusionsThe fast protocol permits dosimetric results within 1 h when IMRT plans are verified, with similar results as those reported by the standard 24 h protocol. 相似文献
5.
PurposeEPID-based in vivo dosimetry (IVD) has been implemented for stereotactic body radiotherapy treatments of non-small cell lung cancer to check both isocenter dose and the treatment reproducibility comparing EPID portal images.Methods15 patients with lung tumors of small dimensions and treated with volumetric modulated arc therapy were enrolled for this initial experience. IVD tests supplied ratios R between in vivo reconstructed and planned isocenter doses. Moreover a γ-like analysis between daily EPID portal images and a reference one, in terms of percentage of points with γ-value smaller than 1, Pγ<1, and mean γ-values, γmean, using a local 3%–3 mm criteria, was adopted to check the treatment reproducibility. Tolerance levels of 5% for R ratio, Pγ<1 higher than 90% and γmean lower than 0.67 were adopted.ResultsA total of 160 EPID images, two images for each therapy session, were acquired during the treatment of the 15 patients. The overall mean of the R ratios was equal to 1.005 ± 0.014 (1 SD), with 96.9% of tests within ± 5%. The 2 D image γ-like analysis showed an overall γmean of 0.39 ± 0.12 with 96.1% of tests within the tolerance level, and an average Pγ<1 value equal to 96.4 ± 3.6% with 95.4% of tests with Pγ<1 > 90%. Paradigmatic discrepancies were observed in three patients: a set-up error and a patient morphological change were identified thanks to CBCT image analysis whereas the third discrepancy was not fully justified.ConclusionsThis procedure can provide improved patient safety as well as a first step to integrate IVD and CBCT dose recalculation. 相似文献
6.
7.
ObjectiveDifferent dose response functions of EBT3 model GafChromic™ film dosimetry system have been compared in terms of sensitivity as well as uncertainty vs. error analysis. We also made an assessment of the necessity of scanning film pieces before and after irradiation.MethodsPieces of EBT3 film model were irradiated to different dose values in Solid Water (SW) phantom. Based on images scanned in both reflection and transmission mode before and after irradiation, twelve different response functions were calculated. For every response function, a reference radiochromic film dosimetry system was established by generating calibration curve and by performing the error vs. uncertainty analysis.ResultsResponse functions using pixel values from the green channel demonstrated the highest sensitivity in both transmission and reflection mode. All functions were successfully fitted with rational functional form, and provided an overall one-sigma uncertainty of better than 2% for doses above 2 Gy. Use of pre-scanned images to calculate response functions resulted in negligible improvement in dose measurement accuracy.ConclusionAlthough reflection scanning mode provides higher sensitivity and could lead to a more widespread use of radiochromic film dosimetry, it has fairly limited dose range and slightly increased uncertainty when compared to transmission scan based response functions. Double-scanning technique, either in transmission or reflection mode, shows negligible improvement in dose accuracy as well as a negligible increase in dose uncertainty. Normalized pixel value of the images scanned in transmission mode shows linear response in a dose range of up to 11 Gy. 相似文献
8.
Pretreatment intensity-modulated radiotherapy quality assurance is performed using simple rectangular or cylindrical phantoms; thus, the dosimetric errors caused by complex patient-specific anatomy are absent in the evaluation objects. In this study, we construct a system for generating patient-specific three-dimensional (3D)-printed phantoms for radiotherapy dosimetry. An anthropomorphic head phantom containing the bone and hollow of the paranasal sinus is scanned by computed tomography (CT). Based on surface rendering data, a patient-specific phantom is formed using a fused-deposition-modeling-based 3D printer, with a polylactic acid filament as the printing material. Radiophotoluminescence glass dosimeters can be inserted in the 3D-printed phantom. The phantom shape, CT value, and absorbed doses are compared between the actual and 3D-printed phantoms. The shape difference between the actual and printed phantoms is less than 1 mm except in the bottom surface region. The average CT value of the infill region in the 3D-printed phantom is −6 ± 18 Hounsfield units (HU) and that of the vertical shell region is 126 ± 18 HU. When the same plans were irradiated, the dose differences were generally less than 2%. These results demonstrate the feasibility of the 3D-printed phantom for artificial in vivo dosimetry in radiotherapy quality assurance. 相似文献
9.
《Reports of Practical Oncology and Radiotherapy》2020,25(1):35-40
AimThis study retrospectively analysed the results of using optically stimulated radiation dosimeters (OSLDs) for in vivo dose measurements during total skin electron therapy (TSET, also known as TSEI, TSEB, TSEBT, TSI or TBE) treatments of patients with mycosis fungoides.BackgroundTSET treatments are generally delivered to standing patients, using treatment plans that are devised using manual dose calculations that require verification via in vivo dosimetry. Despite the increasing use of OSLDs for radiation dosimetry, there is minimal published guidance on the use of OSLDs for TSET verification.Materials and methodsThis study retrospectively reviewed in vivo dose measurements made during treatments of nine consecutive TSET patients, treated between 2013 and 2018. Landauer nanoDot OSLDs were used to measure the skin dose at reference locations on each patient, as well as at locations of clinical interest such as the head, hands, feet, axilla and groin.Results1301 OSLD measurements were aggregated and analysed, producing results that were in broad agreement with previous TLD studies, while providing additional information about the variation of dose across concave surfaces and potentially guiding future refinement of treatment setup. In many cases these in vivo measurements were used to identify deviations from the planned dose in reference locations and to identify anatomical regions where additional shielding or boost treatments were required.ConclusionsOSLDs can be used to obtain measurements of TSET dose that can inform monitor unit adjustments and identify regions of under and over dosage, while potentially informing continuous quality improvement in TSET treatment delivery. 相似文献
10.
Gabriele Guidi Giovanni Gottardi Paola Ceroni Tiziana Costi 《Reports of Practical Oncology and Radiotherapy》2014,19(2):144-150
This work reviews results of in vivo dosimetry (IVD) for total skin electron beam (TSEB) therapy, focusing on new methods, data emerged within 2012. All quoted data are based on a careful review of the literature reporting IVD results for patients treated by means of TSEB therapy. Many of the reviewed papers refer mainly to now old studies and/or old guidelines and recommendations (by IAEA, AAPM and EORTC), because (due to intrinsic rareness of TSEB-treated pathologies) only a limited number of works and reports with a large set of numerical data and proper statistical analysis is up-to-day available in scientific literature. Nonetheless, a general summary of the results obtained by the now numerous IVD techniques available is reported; innovative devices and methods, together with areas of possible further and possibly multicenter investigations for TSEB therapies are highlighted. 相似文献
11.
PurposeTo study how noise and scanner response variability affect radiochromic film dosimetry.MethodsFive treatment plans were analyzed in this work with two different multichannel protocols: the multichannel algorithm of Mayer et al. and the efficient protocol of Lewis et al.Results and conclusionThe multichannel protocol of Mayer et al. is not able to compensate variability in scanner response, which is an important issue for radiochromic film dosimetry. The efficient protocol compensates variations of scanner response, so dose values and gamma scores become more accurate and reproducible. The compensation of digitizer scan variability of the efficient protocol, together with time averaging improve radiochromic film dosimetry. Noise is related to selected resolution in the scanner, our results show that if high resolution measurements are required, de-noising should be considered. 相似文献
12.
PurposeTo analyze the changes in the polarization state of the flatbed scanner light caused by the EBT3 films and to propose a new method for correcting the lateral effects.Methods and materialsThe polarization changes induced by radiochromic films are analyzed using linear polarizing film. Based on the results, the linear polarizing films are used in the scanning process of the EBT3 films. This method is tested against the conventional EBT3 dosimetry using a series of simple regular beams and 21 cases of IMRT.ResultsThe mean results are statically different from the conventional dosimetry with EBT3. Depending on the transmission axis of the polarizing sheet, the results are better or worse compared to conventional dosimetry EBT3 film. When the transmission axis of the polarizing sheet is parallel to the coating direction, the dosimetry results are better and its variability is smaller. However, when the polarizer transmission axis is perpendicular to the coating direction, results are worse as well as its variability.ConclusionUsing a polarized film with the polarization axis parallel to the coating direction of the radiochromic film, and preferably above it, significantly improves the dosimetry results and is an easy and inexpensive way to correct the lateral artifacts of the conventional EBT3 dosimetry. 相似文献
13.
PurposeThis study aims at characterising the properties of TruView™ and ClearView™ two new gel dosimeters (Modus Medical Devices Inc.) and at studying the feasibility of relative dosimetry using these dosimeters and the Vista™ Optical CT scanner to accurately evaluate dose.MethodsIn this work, we investigated key dosimetric aspects (dose response, energy and dose rate dependence) and stability of these radiochromic gels initiated in preliminary works (Huet et al., 2017; Colnot et al., 2017) using spectrophotometric measurements. Moreover, by mean of optical CT scanning (Vista™), their performances to measure relative depth dose (PDD) and cross profiles were analysed.ResultsTruView™ and ClearView™ present a linear dose response up to 20 Gy and up to 80 Gy respectively, independent of both photon beam energy (4–18 MV) and dose rate (up to 9.9 Gy/min) (Huet et al., 2017; Colnot et al., 2017). ClearView™ response proves to be stable for a week post-irradiation and uniform within the batch whereas TruView™ presents an unstable but uniform response. Optical CT scanning generates errors due to stray light that need to be corrected in order to use these gels; ClearView™ scanning particularly requires important precautions. After corrections, those gels used in combination with the Vista™ scanner show promising spatial and dosimetric precision (dose difference <5%). Finally, TruView™ is reusable and presents excellent reproducible response (maximum 3% difference) and the ClearView™ dosimeter presents good spatial stability (0.5% difference after 6 days).ConclusionThis study provides important knowledge about two gel dosimeters presenting interesting dosimetric properties. A study is ongoing to benchmark those promising candidates for clinical dose verification. 相似文献
14.
Georgina Fröhlich Kinga Dóra Kovács Tibor Major Csaba Polgár 《Reports of Practical Oncology and Radiotherapy》2019,24(2):158-164
Aim and backgroundTo investigate the feasibility of in vivo rectal dosimetry in image-guided adaptive brachytherapy of cervical cancer.Materials and methodsError of measurement of dose rate in a semiconductor diode probe was investigated depending on the distance and angle in water, and on temperature in a polymethyl methacrylate phantom using an Ir-192 source. Furthermore, the difference between the measured and calculated dose was analysed in the interstitial brachytherapy of 30 cervix cancer patients. The relationship between in vivo measured dose, calculated dose in the point of the diode, calculated maximal dose in the point of the diodes and calculated maximal dose of the rectum were examined.ResultsThe dosimeter measured with 85% accuracy at more than 5 cm from the source, but within a closer distance the accuracy decreased significantly. At 45–90° angle, the device measured with a 15% error. The error increased with the temperature, 22% at 35 °C. In 8 cases (26.7%) the maximal dose was measured in the correct diode. The device measured 73% of the calculated dose in the point of the diode. The maximum of the calculated doses of diodes was 60% of the calculated maximal dose. The in vivo measured dose was 35% of the calculated maximal dose.ConclusionsUnder treatment conditions, the semiconductor diode does not provide reliable measured data. The probe pushes the rectal wall closer to the high dose areas and underestimates the dose of it. Semiconductor probe is not recommended for in vivo dosimetry of the rectum in image-guided brachytherapy of cervical cancer. 相似文献
15.
Dorota Maria Borowicz Julian Malicki Gennady Mytsin Konstantin Shipulin 《Reports of Practical Oncology and Radiotherapy》2014,19(1):12-18
Aim
To present the results obtained using radiochromic films EBT and RTQA 1010P for the reconstruction the dose distributions for targets irradiated by proton beam and modified by wax boluses.Background
In Medico-Technical Complex at the Joint Institute for Nuclear Research in Dubna implemented technology of wax boluses.Materials and methods
Wax boluses are easier to make and they give better dose distributions than boluses made from modeling clay previously used at our center. We irradiated two imaginary targets, one shaped as a cylinder and the other one as two cuboids. The evaluated calibration curve was used for calculation of the dose distributions measured by the EBT and RTQA radiochromic film. In both cases, the measured dose distributions were compared to the dose distributions calculated by the treatment planning system (TPS). We also compared dose distributions using three different conformity indices at a 95% isodose.Results
Better target coverage and better compliance of measurements (semiconductor detectors and radiochromic films) with calculated doses was obtained for cylindrical target than for cuboidal target. The 95% isodose covered well the tumor for both target shapes, while for cuboidal target larger volume around the target received therapeutic dose, due to the complicated target shape. The use wax boluses provided to be effective tool in modifying proton beam to achieve appropriate shape of isodose distribution.Conclusion
EBT film yielded the best visual matching. Both EBT and RTQA films confirmed good conformity between calculated and measured doses, thus confirming that wax boluses used to modify the proton beam resulted in good dose distributions. 相似文献16.
PurposeThis study provides methodology of calibrating as well as controlling the output for an MTT (3-(4,5-dimethylthiazol-2-yl)-2,5-diphenyltetrazolium bromide) colorimetric assay irradiated in a low energy proton beam using EBT3-model GAFCHROMICTM film, without correcting for quenching effect.MethodsA calibrated Markus ionization chamber was used to measure the depth dose and beam output for 26.5 MeV protons produced by a CS30 cyclotron. A time-controlled aluminum cylinder was added in front of the horizontal beam-exit serving as a radiation shutter. Following the TRS-398 reference dosimetry protocol for proton beams, the output was calibrated in water at a reference depth of 3 mm. EBT3 film was calibrated for doses up to 8 Gy at the same depth. To verify the dose distribution for each 96-well MTT assay plate, EBT3 film was placed at the reference depth during irradiation and cell doses were scaled by measured percent depth dose (PDD) data.ResultsThe radiochromic film dosimetry system in this study provides dose measurements with an uncertainty better than 3.3% for doses higher than 1 Gy. From a single exposure and utilizing the Gaussian shape of the beam, multiple dose points can be obtained within different wells of the same plate ranging from 6.9 Gy (sigma ∼4%) in the central well, and 2 Gy (sigma ∼8%) for wells positioned closer to the periphery.ConclusionsWe described a methodology for radiochromic film-based dose monitoring system, using low-energy protons, which can be used for the MTT assay in any proton beam, except within Bragg peak region. 相似文献
17.
PurposePhotographic film is widely used for the dose distribution verification of intensity-modulated radiation therapy (IMRT). However, analysis for verification of the results is subjective. We present a novel method for marking the isocenter using irradiation from a megavoltage (MV) beam transmitted through slits in a multi-leaf collimator (MLC).MethodsWe evaluated the effect of the marking irradiation at 500 monitor units (MU) on the total transmission through the MLC using an ionization chamber and Radiochromic Film. Film dosimetry was performed for quality assurance (QA) of IMRT plans. Three methods of registration were used for each film: marking by irradiating with an MV beam through slits in the MLC (MLC-IC); marking with a fabricated phantom (Phantom-IC); and a subjective method based on isodose lines (Manual). Each method was subjected to local γ-analysis.ResultsThe effect of the marking irradiation on the total transmission was 0.16%, as measured by a ionization chamber at a 10-cm depth in a solid phantom, while the inter-leaf transmission was 0.3%, determined from the film. The mean pass rates for each registration method agreed within ±1% when the criteria used were a distance-to-agreement (DTA) of 3 mm and a dose difference (DD) of 3%. For DTA/DD criteria of 2 mm/3%, the pass rates in the sagittal plane were 96.09 ± 0.631% (MLC-IC), 96.27 ± 0.399% (Phantom-IC), and 95.62 ± 0.988% (Manual).ConclusionThe present method is a versatile and useful method of improving the objectivity of film dosimetry for IMRT QA. 相似文献
18.
Nahoko Ishioka Shigeko Umeki Yuko Hirai Mitoshi Akiyama Takashi Kodama Kozo Ohama Seishi Kyoizumi 《Mutation Research - Genetic Toxicology and Environmental Mutagenesis》1997,390(3):269
The T-cell receptor (TCR) mutation assay for in vivo somatic mutations is a sensitive indicator of exposure to ionizing radiation. However, this assay cannot be immediately applied after radiation exposure because expression of a mutant phenotype may require as long as several months. In the present study, we eliminate this time lag by stimulating lymphocytes with a mitogen that can accelerate the turnover of TCR protein expression in T-cells. When lymphocytes obtained from healthy donors were irradiated with various doses of X-rays and cultured with human interleukin-2 after phytohemagglutinin (PHA) pulse stimulation, the mutant frequency (MF) of CD4+ T-cells increased dose dependently during the first 7 days, then decreased rapidly due to the growth disadvantage of mutant cells. This suggests that PHA stimulation can shorten the expression time of a mutant phenotype to within a week after radiation exposure. The relationship between radiation dose and TCR MF on the seventh day was best fitted by a linear-quadratic dose–response model. We applied this improved TCR mutation assay to gynecological cancer patients who received 5 days of localized radiotherapy, totaling about 10 Gy. The in vivo TCR MF in the patients did not change within a week after radiotherapy, whereas the in vitro TCR MF of PHA-stimulated lymphocytes from the same patients significantly increased 7 days after initiating culture. The estimated mean radiation dose to the peripheral blood lymphocytes of the cancer patients was about 0.9 Gy, based on the in vitro linear-quadratic dose–response curve. This estimated dose was close to that described in a previous report on unstable-type chromosome aberrations from cervical cancer patients after receiving the same course of radiotherapy. On the basis of these findings, we propose that the improved TCR mutation assay is a useful biological dosimeter for recent radiation exposure. 相似文献
19.
20.
PurposePost-implant CT-scanning is an essential part of permanent prostate brachytherapy. However, the evaluation of post-implant CT dosimetry is not straightforward due to the edema that can modify the dose to the prostate and to the organs at risk. The aim of this study is to evaluate the impact of the timing of the post-implant CT-scan on the dosimetric results and to verify if the Day 0 scan findings can predict Day 50 scanning.Methods136 consecutive patients who received monotherapy with I-125 implants were selected for this study. Two sets of 8 dosimetric quality parameters corresponding to 2 different CT-scans (Day 0 and Day 50) were calculated and compared. The dosimetric parameters included are the percentage volume of the post-implant prostate receiving 80%, 100% and 150% of the prescribed dose, the doses covering 80% and 90% of the prostate volume and the Dose Homogeneity Index. The values of the dose covering 1 cm3 of the rectum and urethra were assessed.ResultsAll the dosimetric parameters of the Day 50 were higher than those of the Day 0 scan. Linear functions were obtained that calculate D90 and V100 values at Day 50 based on the Day 0 findings. Rectal and urethral parameters tended to be underestimated on Day 0 CT-scan relative to Day 50 based dosimetry.ConclusionsPredicting the Day 50 dosimetry from the Day 0 scan could be a possible alternative to a Day 50 scan only in specific situations, but with a degree of uncertainty in the predicted values. 相似文献